Trouble
Cause
Remedy
9. Objective hits the specimen when
The specimen is upside down
Set the specimen correctly with the
an objective is switched to a higher
cover glass on the top
magnification objective
The cover glass is too thick
Use a cover glass with thickness of
0.17mm
The stage is raised too high
Lower the stage
The slide is slipped from the slide
Re-position the slide in the slide
holder
holder
Slide is of excessive thickness
Use slide having thickness between
0.9 and 1.4mm
10. LED does not light
LED is not mounted
Attach LED
LED is blown
Replace the LED
The power cord is unplugged
Plug it securely
Fuse is blown
Check and replace with live fuse
Battery is low
Charge battery
11. LED blows easily
The specified LED is not used
Replace with a specified LED