INSTALLATION OF MICROSCOPE (MOBILE STAND)
6129000-795
Issue 1.7
Printed on May,2020
J
K
L
M
N
O
5
L.1
6
P
36
Fig. 10
Fig. 9
4. Install the standard accessories onto the carrier assembly as
mentioned below:
a. Loosen the M4 grub screw (3) by 2mm allen key and Fix Double
iris diaphragm (J) on magnichanger by re-tightening the m4 grub
screw.
b. Loosen the M4 grub screw (4) by 2mm allen key on Double Iris
Diaphragm and fix DBSi (Double beam splitter inclined) (K) and
re-tighten the M4 grub screw.
c. Remove the End cap from DBSi by rotating the ring (5) in Anticlock-
wise and fix DSLR Camera adapter (L) by matching the slot and
rotate the ring (5) clockwise to tighten the same. Loosen the knob
(6) on DSLR adapter and fix the supplied Mount (Canon, Nikon,
Sony (L.1). Retighten the Knob (6) to secure C-mount.
d. Loosen the M4 grub screw in DBSi by 2mm allen key and fix the
Rotoplate (M). Re tighten the M4 grub screw.
e. Loosen the M4 grub screw by 2mm allen key on Rotoplate and
Fix Ergo Head (0˚ - 210˚ tiltable) (N) and retighten the M4 grub
screw.
f. Fix both eyepieces
(O)
on ergo head.
Note: Please ensure the accessories were fixed properly to have
better visual results.
5. Mount the light guide by completing the steps below:
a. Switch off the illumination system.
The light guide comes pre-routed through the swivel arm (which
houses the LED illumination system, the suspension arm and
through the cable cover.
b. A wire mesh for Cable routing is Permanently fixed to carrier arm
at point (P).
c. Fold and gather wire mesh at Point (P) with one hand and Insert
the Fibre light guide into mesh with other hand.
d. Now release the wire mesh and Pasten its other end to cable cover
at point (H). Referring Fig.14.
e. Make sure the light guide has been routed through wire mesh into
the carrier system such that the microscope is not obstructed and
can be moved in its entire range of movement without stretching,
Kinking, or twisting the light guide.
4
3
Fig. 9