CARE AND MAINTENANCE
6129000-795
Issue 1.7
Printed on May,2020
49
CARE INSTRUCTIONS
1. Keep accessories away from dust when not in use, e.g. protect them with dust cover.
2. Remove dust with a pneumatic rubber blower and soft brush.
3. Protect your microscope from moisture, fumes, acids and cosmetic materials. Do not store chemical
close to the instrument.
4. Protect it from improper handling. Never install other devices sockets or unscrew optical system
and mechanical parts unless explicitly instructed to do so in this manual.
5. Protect the microscope from oil and grease. Never oil or grease the guide surfaces or mechanical
parts.
6. Remove coarse contamination using a damp disposable cloth.
7. Use disinfectants based on the following ingredients: aldehyde's, alcohols, quaternary ammonium
compounds.
8. Use special optics cleaning cloths and pure alcohol for cleaning lenses and eyepieces.
9. Clean optical components using a lint-free cloth. Soak the cloth using a little methanol or Glass cleaner.
Do not use ethanol and spirit.
10. Do not clean products and optical components in a cleaning/disinfecting devices or ultra sound bath.
11.
LABOMED Max Lite
coating are fungal resistant. If you clean as described above, the coatings
will not be damaged.
12. If cleaning and disinfection is done as described above there is no impact of multiple cleaning/
Disinfection on mechanical and optical parts. Paints & Coating are rigid and life long.
TROPICAL ENVIRONMENT / FUNGUS
LABOMED
employs certain safety precaution in its manufacturing techniques and materials. Other preven-
tive measures include:
•
Keep optical parts clean.
•
Use and store them in a clean, cool and dry environment only.
•
Keep moisture away using silica gel and dust cover.
OCCUPATIONAL SAFETY AND HEALTH PROTECTION
•
Observe work safety and health protection of persons responsible for processing contaminated products.
•
Current regulations of hospital hygiene and prevention of infection must be observed in the prepara-
tion, cleaning and disinfection of the products.
INSTRUCTIONS WORKPLACE
•
Remove surface contamination with a paper towel.
AUTOCLAVING
The rubber caps, sleeves and grips supplied by labomed are recommend for the following program for
autoclaving:
•
Temperature: 134˚C
•
Time: 10 minutes
•
Instrument: Standard Autoclave