BA_072_2020-02_10.00  Pressure transmitter PASCAL Ci4 

Page 4/28 

3.4  Mounting the Remote Display 

Optionally you can mount the display and control unit away from the measuring point in an 

additional housing. The back plate of the housing is universally suitable for wall mounting 

or pipe mounting for pipe diameters from 30 – 64 mm. For pipe mounting, you can order 

the corresponding U-bolts from LABOM. 
For best EMC protection, only use the included cable. 
If you ordered the remote display together with the device, the assembly has been com-

pleted in the factory. You only need to route the cable and mount the housing with the re-

mote display. 
If you have received the remote display as a retrofit kit, proceed as follows: 

  

1.  Unmount the front cover and the display unit from the PASCAL Ci4 (see 4.2). 
2.  Replace  them  with  the  adapter  piece  (round  part  with  M12  socket)  and  front  cover, 

which  were  screwed  onto  the  remote  display  housing  for  shipment.  Do  not  forget  to 

connect the cable on the back with the display connector on the CPU module. 

3.  The  cable may  be  shortened  to on-site  requirements.  Colour  codes  on  the terminals 

indicate which wire should be connected to which terminal (see Figure 2). 

4.  Mount the remote display housing using the mounting holes to a wall, a pipe or another 

suitable location. 

5.  Plug the M12 connector of the cable into the socket of the adapter piece. 
6.  Now connect the cable on the display module with the corresponding connector on the 

terminal board in the remote housing and insert the display module in the same man-

ner as for the device itself (see 4.2).  

7.  The last step is to close the remote housing with the display cover that was previously 

on the device. 

  

 

 

Colour codes

 

BN/BR: Brown 
WH/WS: White 
BU/BL: Blue 
BK/SW: Black 

  

 

Figure 2: Connection of the remote display housing 

  

This  modification  can  be  performed  during  operation.  We  do  recommend,  however,  to 

switch off the device during the modification. 

  

Содержание CI4 Series

Страница 1: ...EU Regulations 2 1 4 ATEX Approval 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Mechanical Installation 2 3 2 Electrical Connection 3 3 3 Devices with Diaphragm Seal 3 3 4 M...

Страница 2: ...the device certifies its compliance with the applicable EU Directives for placing products on the market within the European Union You find the complete EU Declaration of Conformity document no KE_04...

Страница 3: ...Pay attention to the following points regarding electrical connections via cable gland and terminal block Do not forget to tighten the cable gland after the electrical connection is finalised Press do...

Страница 4: ...and front cover which were screwed onto the remote display housing for shipment Do not forget to connect the cable on the back with the display connector on the CPU module 3 The cable may be shortened...

Страница 5: ...nals on the terminal board When you connect a current meter to the Test and Test terminals the current is automatically routed through it Figure 4 Current measurement during operation using test termi...

Страница 6: ...y contrast can be adjusted display software version 2 0 5 and above Press and hold the ESC key and use the arrow keys to adapt the contrast The setting is saved and the contrast screen closes as soon...

Страница 7: ...Warning Opening pressurized lines might cause severe injuries Danger of severe injuries or damage Relieve the process pressure before attempting to remove the device Shut off the pressure supply for...

Страница 8: ...n of the device The gen eral functionality of the buttons is identical in all operating modes Figure 7 Control elements If you press and hold the up or down button it will automatically be triggered m...

Страница 9: ...ly measured value is shown If the measured value is larger than the displayable number due to setting a fixed decimal point or an inappropriate unit the largest displayable number is shown flashing In...

Страница 10: ...bination to unlock the menue again The menue cannot be un locked with the key combination if the hardware DIP switch is activated 6 2 3 Icons indicating device status In the top right corner of the di...

Страница 11: ...i mum size best readability from a dis tance One additional value can also be shown Table 4 Display layouts You can define what information shall be shown in the layout s placeholders You can select t...

Страница 12: ...value display by pressing and holding the ESC button for at least one second Figure 9 The Operating menue In long menues and selection lists a scroll bar on the left side shows the position of the cu...

Страница 13: ...is not saved With the OK button the selected value is saved Saving the value is confirmed with an in formation window and the new value is shown in view mode With ESC you leave the view mode and retu...

Страница 14: ...rameter limit is reached The value is then set to the parameter limit You can set negative numbers by decreasing the value below zero When you press the ESC button you can cancel the entry at any time...

Страница 15: ...ion System Device data and function such as factory reset Tabelle 6 The main menue 6 4 2 Quick Setup menue In Quick Setup basic configuration options are combined to make it possible to quickly config...

Страница 16: ...ero point correction requires a non pressurized pressure port e g during calibration in the lab When executing this function the applied pressure will be interpreted as zero bar relative pressure The...

Страница 17: ...ent should be performed as close as possible to the lower range value and upper range value of the measurement ranges 6 4 3 4 Current adjustment You can use the current adjustment if the reading at th...

Страница 18: ...en mode Configuration of the screen layout and content see 6 2 4 Decimal point Select setting of decimal point to determine number of decimal places Backlight Switch backlight on off Table 10 Display...

Страница 19: ...si Pound force per square inch 1 psi 68 9 mbar atm Atmospheric pressure 1 atm 1013 mbar mmH2O Millimetre water column 1 mmH2O 0 0981 mbar mH2O Metre water column 1 mH2O 98 1 mbar inH2O Inch water colu...

Страница 20: ...ent output menue Menue entry Description Damping Setting the damping of the output signal Measuring rate Setting the measuring rate 20 or 100 Hz except for devices with differential pressure cell type...

Страница 21: ...pressure value 6 4 5 2 Setting of the measuring rate When shipped the transmitter is set to a measuring rate of 20 Hz i e 20 times per second the pressure is measured and the current is calculated and...

Страница 22: ...tion of the table function With the table function you can realise any output function E g you can convert the fill height to the fill quantity depending on the tank shape You can use up to 64 interpo...

Страница 23: ...the lower limit between 3 8 and 4 mA and for the upper limit be tween 20 and 21 mA 6 4 6 Diagnosis menue In this menue you can view and configure various diagnostic information The following diagnost...

Страница 24: ...ainte nance need is indicated by an icon in the display see 6 2 3 and if necessary the HART protocol When the maintenance timer has expired the counter continues into the minus range thus the degree t...

Страница 25: ...3 will not automatically activate the constant current mode This must be done under the menue Current mode see 6 4 8 2 6 4 8 2 Current mode The current mode determines whether the output current of th...

Страница 26: ...display module You can transfer the data back to the device or to another device Transferring a configu ration to another device requires a compatible target device The same nominal range is not neces...

Страница 27: ...BA_072_2020 02_10 00 Pressure transmitter PASCAL Ci4 Page 27 28...

Страница 28: ...BA_072_2020 02_10 00 Pressure transmitter PASCAL Ci4 Page 28 28 6 4 10 1 1 1Overview with menue tree and device functions...

Отзывы: