BA_036_2019-11_10.01  High temperature pressure gauges 

Page 4/4 

Disassembly 

When measuring hot media, make sure that the device has cooled down prior to any dis-
mounting or wear appropriate protective clothing to avoid burns. 

  

Switch off the power supply to the device before disconnecting the electrical connections. 
Once this is done, the device may be mechanically removed. 

  

Remove the four flange screws. 

The pressure transmitter system may be pulled out only while the process medium is liquid 
(above melting point). 

Use cylinder head screws (M12 x 55 as per DIN 912, A2-70) quality) for the two threaded 
holes, as shown on the drilling diagram, to push the device out of the process and ensure 
ease of removal. You can order these screws as an accessory. 

  

 

Warning 

Opening pressurized lines might cause severe injuries. 

Danger of severe injuries or damage 

  Relieve  the  process  pressure  before  attempting  to  remove  the  device. 

Shut off the pressure supply for all feed lines to the device and relieve the 
pressure in them. 

  

 

Warning 

Hazardous deposits and residues might remain on opened process connec-
tions and removed devices. 

Danger of injury 

  After  the  device  has  been  removed,  seal  off  the  measuring  point  and 

mark the open process connection accordingly. Consider a possible dan-
ger due to residues when handling the removed device. 

  

Содержание BX1007 Series

Страница 1: ...9 4408 804 444 Fax 49 4408 804 100 e mail sales labom com www labom com Page 1 4 Operating Instructions 1 General Information 2 1 1 General Safety Notes 2 1 2 Intended Use 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Mechanical Installation 2 4 Operation 3 4 1 Zero point correction 3 4 2 Maintenance Service 3 5 Disassembly 4 ...

Страница 2: ...he origi nal packaging Avoid exposure to shocks and excessive vibrations Permissible storage temperature 0 70 C 3 Installation and Commissioning Ensure that the device is suitable for the intended application with respect to pressure range overpressure limit media compatibility temperature range and process connection 3 1 Mechanical Installation Use gaskets if required that are suitable for the pr...

Страница 3: ...ambient temperature 10 70 C 4 1 Zero point correction Small measuring errors or deviations caused by difference in level between pres sure gauge and point of measurement can be corrected on measuring devices with micro control position pointers To do so turn the adjusting screw on the pointer hub see figure 4 Similarly a displacement of the zero point caused by use and long service life can be cor...

Страница 4: ... 912 A2 70 quality for the two threaded holes as shown on the drilling diagram to push the device out of the process and ensure ease of removal You can order these screws as an accessory Warning Opening pressurized lines might cause severe injuries Danger of severe injuries or damage Relieve the process pressure before attempting to remove the device Shut off the pressure supply for all feed lines...

Отзывы: