 5.0  Safety  Information

NOTE: 

Be careful when changing the microcentrifuge tube platform, especially when you have used the unit at 

temperatures higher than 60°C. Always wear protective clothing before you handle a hot platform.

The unit will continue to shake for 5 seconds after the lid is opened. Be careful when opening the lid as the parts inside 
may still be in motion. In addition, never touch the fan unless the unit is turned off or unplugged.

 



If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

 



Before cleaning the housing unplug the unit. The housing should only be cleaned with a damp cloth and if necessary, a mild soap. 
Don’t use aggressive or aerosol cleaners.

 



Do not use the unit near sources of water. Take care to ensure that water will not spill in the device, especially during cleaning 
procedures.

 



Make sure, that all test tubes are closed tightly before placing them into the unit to avoid spilling samples inside the chamber.

In the case of a malfunction, unplug the device and contact your distributor for service.

Do not shake flammable or explosive samples.

 6.0  Installation

6.1. Device Placement

Do not use the device in a flammable or explosive atmosphere.

NOTE:

 There has to be easy access to the incubator control panel and main plug in case of emergency.

When selecting the right place for device, please consider following:

 



Put the device on a smooth, horizontal, and stable place.

 



Leave at least 10 cm of space around the device for adequate air circulation.

 



Leave enough space around the device, for easy access and maintenance.

 



Do not place the device where there are rapid temperature and humidity changes. Also avoid places where the unit would be exposed to 
direct sunlight or next to devices that output large amounts of heat.

 



Avoid locations where the unit may be exposed to excessive shocks or vibrations.

 



Avoid locations subject to frequent power fluctuations or power losses.

6.2. Attaching the Power Cord

First, connect the main power cord to the incubator, then connect the plug end of the cord to a grounded wall socket. Avoid lines on 
which powerful electric motors, refrigerators, and similar devices are connected. 

 



Always use caution when plugging the cord into a grounded power outlet.

 



Do not touch the plug with wet hands.

 



Do not pull the plug by the cord.

6.3. Start Up Check List

 



Unpack and install the device as specified in Sections 6.1 and 6.2.

 



Open the lid and attach the microcentrifuge tube platform or the microplate holder platform.

 



Close the lid.

 



Switch on the device using of the ON/OFF switch on the front panel. The switch will illuminate to indicate that the power is On.

 



Check the running parameters, and set new parameters if necessary. Refer to the instructions described in Section 7.

Содержание VorTemp 56

Страница 1: ...Labnet VorTemp 56 Shaking Incubator Instruction Manual Catalog Numbers S2056A S2056A 230V...

Страница 2: ...orm 3 5 0 Safety Information 4 6 0 Installation 4 6 1 Device Placement 4 6 2 Attaching the Power Cord 4 6 3 Start Up Check List 4 7 0 Operating Instructions 5 7 1 Control Panel 5 7 2 Basic Operation 5...

Страница 3: ...Electrical S2056A S2056A 230V 120V 10 50 60Hz 4 8A 230V 10 50 60Hz 2 4A Fuse S2056A S2056A 230V 2 x T 6 3A L 250V 2 x T 3 15A L 250V Environmental Temperature From 5 C to 40 C Relative Humidity Up to...

Страница 4: ...temperature sensor and the control electronics regulate heater function to maintain the set temperature 4 0 Accessories Cat No Description Qty Pk S2056 R 1 5 2 0 mL platform workstation for microcentr...

Страница 5: ...stacked The working temperature range is ambient from 5 C to 40 C Please note that the normal maximum temperature for ordinary microplates is below 60 C The platform is made of stainless steel and is...

Страница 6: ...ss to the incubator control panel and main plug in case of emergency When selecting the right place for device please consider following Put the device on a smooth horizontal and stable place Leave at...

Страница 7: ...ture 6 Encoder Knob Rotate the encoder right or left to modify the Time Temperature or RPM settings Push in the Encoder knob to change between Time Temperature and RPM set values If you rotate the Enc...

Страница 8: ...Temp C Start RPM Illuminated Rotating the encoder right or left sets the rotating speed to the desired value 34 340 rpm 90 900 rpm By turning the RPM setting below 20 the shaking function can be turn...

Страница 9: ...left until the new value is set While adjusting the RPM the light will pulse When you stop rotating the encoder knob the RPM light will stop blinking after 2 sec l l l l l l l l l 37 0 Stop Encoder Po...

Страница 10: ...with the motor regulation PWM regulator pulse generator motor The unit will automatically stop Call for service E 1 2 The motor does not reach the set RPM in 30 sec The unit will automatically stop Ca...

Страница 11: ...chase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be...

Страница 12: ...SIA PACIFIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 India t 91 124 4604000 Japan t 81 3 3586 1996 Korea t 82 2 796 9500 Singapore t 65 6572 9740 Taiwan t 886 2 2716 033...

Отзывы: