|        LABNET VORTEMP 56 SHAKING INCUBATOR

7.5. Additional Operations

Viewing the set value for RPM during shaking

 



Set the unit to RPM mode

 



Turn the encoder ONE CLICK right (+) or left (-). On the display the RPM light will blink for 2 sec. and will display the set RPM. After 
2 sec. the display will show the actual RPM and the light will stop blinking.

Viewing the set value for Time during shaking

 



Set the unit to Time Mode

 



Turn the encoder ONE CLICK right (+) or left (-). The Time light will blink for 2 sec. and will display the set Time. After 2 sec. the 
display will show the actual Time and the light will stop blinking.

Viewing the set value for Temperature during shaking

 



Set the unit to Temperature Mode

 



Turn the encoder ONE CLICK right (+) or left (-). The Temperature light will blink for 2 sec. and will display the set Temperature. After 
2 sec. the display will show the actual Temperature and the light will stop blinking.

 8.0  Troubleshooting

The Labnet VorTemp™ 56 shaking incubator features built in self-diagnostic procedures which are constantly checking the operating 
parameters and performance, as well as functions that are necessary for safe and reliable operation. An error code is shown on the 
LCD display if an error occurs.

Sample of Error display:

Er5

Stop

Encoder

Power

Run

Time [min]

RPM [x10]

Temp [°C]

Start

E 1.1

Something is wrong with the motor regulation (PWM regulator, pulse generator, motor). The unit will automatically stop. 
Call for service.

E 1.2

The motor does not reach the set RPM in 30 sec. The unit will automatically stop. Call for service.

E 1.3

The set RPM oscillates more than 100 RPM in 2 seconds. The unit will automatically stop. Call for service.

E 2.1

The temperature sensor is not working properly with regards to the heater. The unit will automatically stop. Call for service.

E 2.2

The temperature sensor registers a 5°C higher temperature than the set temperature. The unit will automatically stop. 
Call for service.

Er 5

The lid of the shaker is opened for more than 5 seconds. The unit will automatically stop. To restart the shaker, close the lid 
and press Start.

When shipping your equipment for service, follow the packing guidelines listed below:

 



Clean the inside of the chamber and platforms according to GLP (Good Laboratory Practices) standards, especially if you have used 
the equipment with hazardous biological or radioactive materials.

 



A written description of the error should accompany the unit.

 



Use the original shipping container and packaging materials if possible.

 9.0  Temperature  Calibration

NOTE:

 Temperature calibration should only be performed by qualified personnel.

The temperature control software allows for user calibration of the temperature settings. First, measure the temperature in the 
middle of the chamber after allowing the temperature to equilibrate for 2 hours. The temperature should be measured with a digital 

Содержание VorTemp 56

Страница 1: ...Labnet VorTemp 56 Shaking Incubator Instruction Manual Catalog Numbers S2056A S2056A 230V...

Страница 2: ...orm 3 5 0 Safety Information 4 6 0 Installation 4 6 1 Device Placement 4 6 2 Attaching the Power Cord 4 6 3 Start Up Check List 4 7 0 Operating Instructions 5 7 1 Control Panel 5 7 2 Basic Operation 5...

Страница 3: ...Electrical S2056A S2056A 230V 120V 10 50 60Hz 4 8A 230V 10 50 60Hz 2 4A Fuse S2056A S2056A 230V 2 x T 6 3A L 250V 2 x T 3 15A L 250V Environmental Temperature From 5 C to 40 C Relative Humidity Up to...

Страница 4: ...temperature sensor and the control electronics regulate heater function to maintain the set temperature 4 0 Accessories Cat No Description Qty Pk S2056 R 1 5 2 0 mL platform workstation for microcentr...

Страница 5: ...stacked The working temperature range is ambient from 5 C to 40 C Please note that the normal maximum temperature for ordinary microplates is below 60 C The platform is made of stainless steel and is...

Страница 6: ...ss to the incubator control panel and main plug in case of emergency When selecting the right place for device please consider following Put the device on a smooth horizontal and stable place Leave at...

Страница 7: ...ture 6 Encoder Knob Rotate the encoder right or left to modify the Time Temperature or RPM settings Push in the Encoder knob to change between Time Temperature and RPM set values If you rotate the Enc...

Страница 8: ...Temp C Start RPM Illuminated Rotating the encoder right or left sets the rotating speed to the desired value 34 340 rpm 90 900 rpm By turning the RPM setting below 20 the shaking function can be turn...

Страница 9: ...left until the new value is set While adjusting the RPM the light will pulse When you stop rotating the encoder knob the RPM light will stop blinking after 2 sec l l l l l l l l l 37 0 Stop Encoder Po...

Страница 10: ...with the motor regulation PWM regulator pulse generator motor The unit will automatically stop Call for service E 1 2 The motor does not reach the set RPM in 30 sec The unit will automatically stop Ca...

Страница 11: ...chase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be...

Страница 12: ...SIA PACIFIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 India t 91 124 4604000 Japan t 81 3 3586 1996 Korea t 82 2 796 9500 Singapore t 65 6572 9740 Taiwan t 886 2 2716 033...

Отзывы: