background image

  Instruction Manual        | 

         1

 1.0  Introduction

The Labnet VorTemp™ 56 shaking incubator is a benchtop shaker and heating chamber for incubating and shaking sensitive samples. 
It replaces two devices, reducing both time and space needed. It is suitable for biochemistry, microbiology, and clinical laboratories in 
which applications require temperature and shaking treatment.

1.1 Symbols and Conventions

CAUTION: 

This symbol refers you to important operating and maintenance (servicing) instructions within the product 

Instruction Manual. Failure to heed this information may present a risk of damage or injury to persons or equipment.

CAUTION:

 To avoid accidental bodily harming or burning be very careful touching the metal parts of the unit. It can be 

very hot after it is used at high temperatures. Allow the metal parts to cool down before handling.

This symbol indicates the presence of a potential hazard which could result in electrical shock. 

 2.0  Packing

The Labnet VorTemp 56 shaking incubator includes the following components:

 



VorTemp 56 shaking incubator (platforms sold separately)

 



Instruction manual

 3.0  Specifications

Power

515W

Heater Power

500W

Motor Power

15W

Electrical 
   S2056A 
   S2056A-230V

 
120V ±10%, 50/60Hz, 4.8A 
230V ±10%, 50/60Hz, 2.4A

Fuse 
   S2056A 
   S2056A-230V

 
2 x T 6.3A L 250V 
2 x T 3.15A L 250V

Environmental Temperature 

From 5°C to 40°C

Relative Humidity

Up to 85% non-condensing

RPM Regulation

Digital load independent, from 100 to 1,200 rpm in 10 rpm steps, 1,400 rpm (only at 60Hz)

Shaker Orbit

3 mm

Temperature Operating Range

From 5°C above room temperature to 99.5°C

Temperature Sensor

PT100

Heat up Time

Approx. 5°C/min.

Temperature Uniformity

±0.5°C

Timer

From 30 sec. to 99 min. 50 sec. in 10 sec. steps, under 10 min. in 1 sec. steps, timer HOLD function

Maximum Capacity

56 test tubes – 1.5, 0.7, 0.5, 0.25 mL

Dimensions (W x D x H)

10 x 13 x 9 in. (26.5 x 32.5 x 22.5 cm)

Weight

24 lbs. (11 kg)

NOTE:

 The supply cord shall be with PE conductor and cross section at least 3 x 1.00 mm

2

 or 3 x AWG 18 (USA/CANADA).

Labnet VorTemp 56 shaking incubator is designed to be safe when operated under the following conditions:

 



Indoor use

 



Altitude up to 2,000 meters

 



Pollution Degree 2

Unit can be connected to external installation with overcurrent protective device max. value 16A or 20A (USA/CANADA). This device 
is intended for use in a basic electromagnetic environment (Basic, class B), and it is in compliance with IEC 61326-1:2020. Device may 
experience a restart during overvoltage events (surges) or voltage interruptions in mains power supply.

Содержание VorTemp 56

Страница 1: ...Labnet VorTemp 56 Shaking Incubator Instruction Manual Catalog Numbers S2056A S2056A 230V...

Страница 2: ...orm 3 5 0 Safety Information 4 6 0 Installation 4 6 1 Device Placement 4 6 2 Attaching the Power Cord 4 6 3 Start Up Check List 4 7 0 Operating Instructions 5 7 1 Control Panel 5 7 2 Basic Operation 5...

Страница 3: ...Electrical S2056A S2056A 230V 120V 10 50 60Hz 4 8A 230V 10 50 60Hz 2 4A Fuse S2056A S2056A 230V 2 x T 6 3A L 250V 2 x T 3 15A L 250V Environmental Temperature From 5 C to 40 C Relative Humidity Up to...

Страница 4: ...temperature sensor and the control electronics regulate heater function to maintain the set temperature 4 0 Accessories Cat No Description Qty Pk S2056 R 1 5 2 0 mL platform workstation for microcentr...

Страница 5: ...stacked The working temperature range is ambient from 5 C to 40 C Please note that the normal maximum temperature for ordinary microplates is below 60 C The platform is made of stainless steel and is...

Страница 6: ...ss to the incubator control panel and main plug in case of emergency When selecting the right place for device please consider following Put the device on a smooth horizontal and stable place Leave at...

Страница 7: ...ture 6 Encoder Knob Rotate the encoder right or left to modify the Time Temperature or RPM settings Push in the Encoder knob to change between Time Temperature and RPM set values If you rotate the Enc...

Страница 8: ...Temp C Start RPM Illuminated Rotating the encoder right or left sets the rotating speed to the desired value 34 340 rpm 90 900 rpm By turning the RPM setting below 20 the shaking function can be turn...

Страница 9: ...left until the new value is set While adjusting the RPM the light will pulse When you stop rotating the encoder knob the RPM light will stop blinking after 2 sec l l l l l l l l l 37 0 Stop Encoder Po...

Страница 10: ...with the motor regulation PWM regulator pulse generator motor The unit will automatically stop Call for service E 1 2 The motor does not reach the set RPM in 30 sec The unit will automatically stop Ca...

Страница 11: ...chase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be...

Страница 12: ...SIA PACIFIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 India t 91 124 4604000 Japan t 81 3 3586 1996 Korea t 82 2 796 9500 Singapore t 65 6572 9740 Taiwan t 886 2 2716 033...

Отзывы: