Removing the needle cartridge from the packaging
Needle cartridges are packaged individually in sterile blister packs.
Do not open the needle cartridge blister pack until just before the treatment
is to be carried out.
Hands must be sterilised and unused disposable gloves must be worn to
remove the needle cartridge.
Needle cartridges out of damaged or already opened blister packs may
no longer be used because it can no longer be guaranteed that the needle
cartridge is sterile.
Inserting the needle cartridge into the handpiece
NOTE: Preventing damage to the handpiece
Ensure that the pigmentation device is turned off before inserting a
needle cartridge into the handpiece.
Applying slight pressure, insert the dark green needle cartridge into the handpiece
opening as far as it will go. The cartridge is now securely in place in the handpiece.
Turn the needle cartridge a quarter of the way clockwise until you feel resistance.
Checking the needle cartridge before use
After inserting the needle cartridge and before beginning each treatment,
check that the cartridge is securely fixed in the handpiece.
Putting down the handpiece
Make sure that the needle cartridge and handpiece do not come into contact
with anything that it is not intended for. This will prevent the needle cartridge
from becoming contaminated and prevent people and objects from coming
into contact with any blood and lymphatic fluid possibly stuck on the cartridge
or handpiece.
Operating information