Refer to the following table if your Vertical Clean Bench fails to
operate properly. If the suggested corrective actions do not solve
your problem, contact Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Class 5 air conditions
are not present inside
the enclosure.
Improper user
techniques for the
enclosure.
See “Certifying the Enclosure” Chapter
3 and “Safety Precautions” Chapter 4
sections in the manual. (Ref. Appendix
D).
External factors are
disrupting the
enclosure airflow
patterns.
See “Location Requirements” Chapter
2, “Certifying the Enclosure” Chapter
3, and “Safety Precautions” Chapter 4
sections of this manual. (Ref. Appendix
D).
Enclosure has
improper downflow
velocity.
Have enclosure certified and check
exhaust system. Check HEPA filters for
loading. Adjust enclosure speed
control. Enclosure should have an
average downflow velocity of 30-65
fpm.
Содержание Purifier 3970200
Страница 15: ...11 Chapter 3 Getting Started Product Service 1 800 522 7658 Figure 3 1 Vertical Clean Bench Installation ...
Страница 58: ...54 Appendix B Dimensions Product Service 1 800 522 7658 Figure B 1 Purifier Vertical Clean Bench ...
Страница 60: ...56 Appendix C Vertical Clean Bench Specifications Product Service 1 800 522 7658 C 1 ...
Страница 61: ...57 Appendix C Vertical Clean Bench Specifications Product Service 1 800 522 7658 C 2 ...
Страница 62: ...58 Appendix C Vertical Clean Bench Specifications Product Service 1 800 522 7658 C 3 ...
Страница 63: ...59 Appendix C Vertical Clean Bench Specifications Product Service 1 800 522 7658 C 4 ...