8
Product Service 1-800-522-7658
Chapter 2: Prerequisites
by a qualified electrician. Refer to
Chapter 3: Getting Started
and
Appendix
C
: Protector XStream Specifications
for the wiring
diagram for proper electrical installation.
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch
outside diameter, copper (brass for natural gas), and equipped with
an easily accessible shut-off valve, should disconnection be
required. If the service line pressure exceeds 40 PSI, it must be
equipped with a pressure regulator to reduce the line pressure.
Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in
Appendix B:
Protector XStream Dimensions
.
Содержание Protector XStream 9840400
Страница 2: ...Product designs are subject to change without notice 2004 Labconco Corporation Printed in U S A...
Страница 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Страница 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Страница 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Страница 48: ...43 Product Service 1 800 522 7658 Appendix A Protector Components 20 16 21 11 3 1 5 6 12 2 10 19 14 13 7 22 23...
Страница 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Страница 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Страница 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 61: ...56 Product Service 1 800 522 7658 Appendix D References...
Страница 62: ...57 Product Service 1 800 522 7658...