7
Product Service 1-800-522-7658
Chapter 2: Prerequisites
Full Open 28" Sash Opening
Hood
Size
Standard Model Description
Face
Velocity
Exhaust
Volume
Static
Pressure
Loss
4'
48" Protector XStream Hood
60 fpm
440 CFM
0.08" H
2
O
80 fpm
590 CFM
0.13" H
2
O
100 fpm 730 CFM
0.25" H
2
O
5'
60" Protector XStream Hood
60 fpm
580 CFM
0.11" H
2
O
80 fpm
770 CFM
0.19" H
2
O
100 fpm 960 CFM
0.30" H
2
O
6'
72" Protector XStream Hood
60 fpm
710 CFM
0.13" H
2
O
80 fpm
940 CFM
0.23" H
2
O
100 fpm 1180 CFM
0.36" H
2
O
8'
96" Protector XStream Hood
60 fpm
1000 CFM
0.10" H
2
O
80 fpm
1330 CFM
0.18" H
2
O
100 fpm 1660 CFM
0.28" H
2
O
60% Open 18" Sash Opening (sash
stop required)
Hood
Size
Standard Model Description
Face
Velocity
Exhaust
Volume
Static
Pressure
Loss
4'
48" Protector XStream Hood
60 fpm
280 CFM
0.04" H
2
O
80 fpm
380 CFM
0.07" H
2
O
100 fpm
470 CFM
0.11" H
2
O
5'
60" Protector XStream Hood
60 fpm
370 CFM
0.05" H
2
O
80 fpm
490 CFM
0.08" H
2
O
100 fpm
610 CFM
0.13" H
2
O
6'
72" Protector XStream Hood
60 fpm
450 CFM
0.06" H
2
O
80 fpm
600 CFM
0.10" H
2
O
100 fpm
750 CFM
0.15" H
2
O
8'
96" Protector XStream Hood
60 fpm
640 CFM
0.04" H
2
O
80 fpm
850 CFM
0.08" H
2
O
100 fpm
1060 CFM
0.12" H
2
O
Proper blower selection can be determined from these exhaust
requirements and the total system static pressure loss. Contact
Labconco Customer Service for assistance in sizing a remote
blower system.
Electrical Requirements
The Protector XStream Hood models feature internal wiring for the
fluorescent light assembly and light switch. All internal wiring is
terminated at the single point wiring junction box for hook-up by a
qualified electrician. The blower switch, and light switch wires are
also terminated at the single point wiring junction box for hook-up
Содержание Protector XStream 9840400
Страница 2: ...Product designs are subject to change without notice 2004 Labconco Corporation Printed in U S A...
Страница 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Страница 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Страница 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Страница 48: ...43 Product Service 1 800 522 7658 Appendix A Protector Components 20 16 21 11 3 1 5 6 12 2 10 19 14 13 7 22 23...
Страница 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Страница 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Страница 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 61: ...56 Product Service 1 800 522 7658 Appendix D References...
Страница 62: ...57 Product Service 1 800 522 7658...