16
Product Service 1-800-522-7658
Chapter 3: Getting Started
through the work surface to make these connections to the service
fixtures.
NOTE: Inspect all fittings for leakage. Tighten
the fittings slightly if needed.
Should access to the hood plumbing fixture bodies be required,
remove the service access plate on the hood front corner posts by
loosening their individual screws. (See item 11 Figure A-1, page
50 and 51) The valve body will now be fully exposed for any
service work that may be necessary. The service fixtures supplied
on your laboratory hood are designed for use with the following
services:
•
Air
•
Hot Water
•
Vacuum
•
Cold Water
•
Natural Gas – See Caution Below
Sealing the Protector XStream
Hood to the Work Surface
When the hood has been set in place, ducted, wired, and plumbed,
it should be sealed at the work surface to prevent spilled materials
from collecting under the walls of the hood. Materials such as
silicone sealants are recommended to seal the hood structure.
!
CAUTION
:
Do not use oxygen with any standard
service fixture. Contact Labconco Customer
Service for oxygen fixture information.
!
WARNING
:
Contact Labconco Customer
Service directly before using any service other
than those listed above in these valves to assure
full compatibility.
!
CAUTION:
Natural gas should be used only in
the service fixture that has been pre-plumbed
with brass tubing. Sulfur content of the gas
could cause deterioration of standard copper
supply lines.
!
Содержание Protector XStream 9840400
Страница 2: ...Product designs are subject to change without notice 2004 Labconco Corporation Printed in U S A...
Страница 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Страница 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Страница 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Страница 48: ...43 Product Service 1 800 522 7658 Appendix A Protector Components 20 16 21 11 3 1 5 6 12 2 10 19 14 13 7 22 23...
Страница 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Страница 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Страница 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Страница 61: ...56 Product Service 1 800 522 7658 Appendix D References...
Страница 62: ...57 Product Service 1 800 522 7658...