Appendix G: Acid/Formaldehyde Sensor Locations
Fan boxes that have acid or formaldehyde sensors will be labeled with small ‘A’ or ‘F’
stickers in the bottom left corner of the fan face and the carton it shipped in. The
installment locations for the sensors will be as follows:
Figure G-1
*
*
*
*
*
Key:
S
= Solvent Sensor (standard on all Fan Modules)
A
= Acid Sensor (optional)
F
= Formaldehyde Sensor (optional)
T
= Temp Sensor (standard on all Base Frames)
H
= Relative Humidity Sensor (standard on all Base Frames)
Notes:
1)
3-foot GFH: This filtered hood supports a max of (2) chemical
sensors. Available configurations are: S only, S + A, or S + F
2)
4-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 2 and Molecode F on
column 1.
3)
5-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 3 and Molecode F on
column 2.
4)
6-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 3 and Molecode F on
column 2.
5)
8-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 4 and Molecode F on
column 3.
*Molecode S cell replacement for maintenance every 5 years
Содержание Protector Echo
Страница 78: ...78 Fans and Lights not working...
Страница 79: ...79 Vertical sash no longer operates smoothly...
Страница 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Страница 81: ...81 Contaminates outside the filtered hood...
Страница 82: ...82 Lights operate but fans will not...
Страница 83: ...83 Smart Command Alarms...
Страница 84: ...84 If needed contact Labconco to troubleshoot further...
Страница 88: ...88 1 2 3 4 5 6 7 8 9 10 13 12 11 14 15 19 22 23 24 26 25 27 28 20 18 REFERENCE SERIAL TAG Figure A 1...
Страница 94: ...94 Wiring Diagram Figure C 1...
Страница 109: ...109 Figure E 5 TYPICAL NEUTRODINE FILTRATION TECHNOLOGY 12 SCREWS CORNER POST FRONT PANEL HEADER...