Chapter 3: Getting Started
Product Service 1-800-522-7658
10
Original instructions
After you verify the fume hood components, move your hood to the location where
you want to install it.
Sash Weight Release
To protect the fume hood from damage in shipment, the sash weight has been
secured to the back of the fume hood with four (4) screws. Simply remove the
screws and make sure the sash cables are on the pulleys before operation of the
sash.
NOTE
: THE SASH WEIGHT ITSELF WAS INDIVIDUALLY
MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT
BE EXCHANGED ON ANY OTHER UNIT.
Install the Fiberglass 30 Hood on a
Supporting Structure and Work Surface
The Fiberglass 30 Hood is heavy! Use caution when lifting or
moving the unit.
La fibre de verre 30 Hood est lourd! Soyez prudent lorsque
vous soulever ou déplacer l'unité
When installing the Fiberglass 30 Fume Hood onto a chemically resistant work
surface or benchtop, ensure that the structure can safely support the combined
weight of the fume hood and any related equipment. The work surface should be
at least as wide as the hood to properly support it.
The work surface is aligned
with the back of the fume hood. This will provide the correct spacing under
the air foil for proper bypass airflow.
WARNING: It is important to support the rear of the work
surface and fume hood. The cross support provides support for
the bottom of the work surface. Install the cross support after
the base cabinets and work surface are leveled and before
installing the hood.
AVERTISSEMENT : Il est important de soutenir l'arrière de la
surface de travail et la hotte aspirante. Le support tranversal
soutient le bas de la surface de travail. Installer le support
transversal après que les meubles et la surface de travail soient
nivelés et avant d'installer la hotte.
The following are instructions for mounting a cross support:
1.
Level the base cabinets and the work surface. Work surface should be
placed flush with the back of the fume hood as shown in Figure 3-1.
2.
Scribe a line on the wall or back of the base cabinet to locate the
support under the work surface.
!
!
Содержание Fiberglass 30 Series
Страница 1: ...Fiberglass 30 Laboratory Fume Hoods Models 3030000 3030001 3030002 3030003 3030004 3030005...
Страница 43: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 39 Original instructions Figure B 2...
Страница 46: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 42 Original instructions Figure B 5...
Страница 48: ...Appendix C Fiberglass 30 Specifications Product Service 1 800 522 7658 44 Original instructions...
Страница 49: ...Appendix C Fiberglass 30 Specifications Product Service 1 800 522 7658 45 Original instructions...