background image

Chapter 2: Prerequisites 

 

Product Service 1-800-522-7658 

Original instructions 

Fiberglass 30 Hood with Explosion Proof Blower, 100 fpm, 310 cfm, capable of 
overcoming 0.12" external static pressure. 

 

Proper blower selection can be determined from these exhaust requirements and 
the total system static pressure loss.  Contact Labconco Customer Service for 
assistance in sizing a blower. 

Electrical Requirements 

Other than explosion proof models, the Fiberglass 30 Hood models feature 
internal wiring for the incandescent light assembly, light switch and blower 
switch.  If your unit has an integral blower, then internal wiring to the blower 
motor and blower switch is included.  Explosion Proof blowers require on site 
wiring.  All internal wiring is terminated at the single point wiring junction box 
for hook-up by a qualified electrician.  Refer to 

Chapter 3: Getting Started 

and 

Appendix 

C

: Fiberglass 30 Specifications

 for the wiring diagram for proper 

electrical installation. 
 

Autre que les modèles d'explosion preuve, les modèles en fibre de verre 30 capot 
disposent câblage interne pour l'ensemble de lumière incandescente, interrupteur 
de lumière et commutateur de soufflerie. Si votre appareil dispose d'une soufflerie 
intégrée, alors le câblage interne du moteur du ventilateur et l'interrupteur du 
ventilateur est inclus. Explosion Proof ventilateurs exigent le câblage du site. Tout 
le câblage interne est terminé à l'câblage boîte de jonction unique point pour le 
branchement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en 
route et à l'Annexe C: fibre de verre 30 Spécifications pour le schéma de câblage 
pour l'installation électrique appropriée.

 

Service Line Requirements 

All service lines to the laboratory fume hood should be ¼ inch outside diameter, 
copper (brass for natural gas), and equipped with an easily accessible shut-off 
valve, should disconnection be required.  Recommended operating pressure is  
40 PSI, with a maximum allowable pressure of 200 PSI.  Consider a pressure 
regulator to reduce line pressure to 40 PSI.  Please check with local codes for 
other requirements. 

Space Requirements 

The dimensions for the different models are shown in 

Appendix B: Fiberglass 30 

Hood Dimensions

Your Next Step 

After you have determined that the location you have selected accommodates the 
installation and operational requirements of your fume hood, you are ready to 
begin installation.  Proceed to 

Chapter 3: Getting Started

 

Содержание Fiberglass 30 Series

Страница 1: ...Fiberglass 30 Laboratory Fume Hoods Models 3030000 3030001 3030002 3030003 3030004 3030005...

Страница 2: ...30 a m and 5 30 p m Central Standard Time Part 2831600 Rev H ECO L354 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable lab...

Страница 3: ...p 7 CHAPTER 3 GETTING STARTED 8 Unpacking Your Fume Hood 8 Removing the Shipping Skid 9 Sash Weight Release 10 Install the Hood on a Supporting Structure Work Surface 10 Connecting to the Hood Exhaust...

Страница 4: ...erations 28 Your Next Step 29 CHAPTER 7 MODIFYING YOUR FIBERGLASS 30 30 Installing Work Surfaces and Storage Cabinets 30 Installing Additional Service Fixtures 30 Installing Guardian Digital Airflow M...

Страница 5: ...manual and keep it handy for future reference If you are unfamiliar with how fume hoods operate please review Chapter 4 Performance Features and Safety Precautions before you begin working in the fum...

Страница 6: ...corrective actions Appendix A Fiberglass 30 Hood Components contains labeled diagrams of all of the components of the fume hoods Appendix B Fiberglass 30 Hood Dimensions contains comprehensive diagram...

Страница 7: ...s par l ic ne de cl plate Ces op rations devraient tre seulement ex cut es par un professionnel agr e L chec pour se conformer aux informations qui suivent une ic ne de cl plate peut r sulter la bles...

Страница 8: ...how laboratory fume hoods operate go to Chapter 4 Performance Features and Safety Precautions For information on the operational characteristics of your laboratory fume hood go to Chapter 5 Using Your...

Страница 9: ...to learn the requirements for your installation site The location requirements The support requirements The exhaust requirements The electrical power requirements The service line requirements The spa...

Страница 10: ...ically resistant work surface See Chapter 7 Modifying Your Fiberglass 30 Hood NE PAS installer la hotte sur un panier chariot ou banc mobile TOUS fibre de verre de 30 installations hotte doit tre perm...

Страница 11: ...cente interrupteur de lumi re et commutateur de soufflerie Si votre appareil dispose d une soufflerie int gr e alors le c blage interne du moteur du ventilateur et l interrupteur du ventilateur est in...

Страница 12: ...ddition to 5 16 3 8 7 16 and 1 2 wrenches ratchets sockets a nut driver set a flat blade screwdriver a Phillips screwdriver and a carpenter level to complete the instructions in the chapter Unpacking...

Страница 13: ...129 LBS 44 59 KG LA PALETTE D EXP DITION PERMET LEVAGE AVEC UN CHARIOT ELEVATEUR MECANIQUE DES SOLS JACK SI VOUS DEVEZ DECROCHER LE HOTTE MANUELLE SUIVRE LES DIRECTIVES SAFE LEVAGE NORMALLY THE FUME...

Страница 14: ...upport the combined weight of the fume hood and any related equipment The work surface should be at least as wide as the hood to properly support it The work surface is aligned with the back of the fu...

Страница 15: ...hing it to the wall or base cabinet 4 Place the hood on top of the work surface and cross support The work surface should be smooth and durable such as a chemically resistant epoxy resin The surface s...

Страница 16: ...elocities away from the hood Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and total system static pressure loss See Chapter 2 for the hood airflo...

Страница 17: ...ique pour la construction en fibre de verre 30 Hoods est n cessaire d inclure la protection de surcharge Un commutateur ou disjoncteur doit tre tout pr s de l quipement et port e facile de l op rateur...

Страница 18: ...blage pour la hotte doit tre effectu e par un lectricien agr et conforme tous les codes lectriques locaux et nationaux Dans la plupart des cas le capot ne n cessiter l utilisation d un conduit blind...

Страница 19: ...re information ATTENTION Ne pas utiliser d oxyg ne avec n importe quel appareil de service standard Contactez Labconco Service la client le pour obtenir des informations de fixation de l oxyg ne Shoul...

Страница 20: ...work and exhaust blower give you the flexibility to change the airflow at the sash opening of your hood To determine the actual face velocity at the sash opening airflow velocity readings will need to...

Страница 21: ...g Started Product Service 1 800 522 7658 17 Original instructions Your Next Step After your fume hood has been installed and certified you are ready to proceed to Chapter 4 Performance Features and Sa...

Страница 22: ...lity of 28 high while conserving energy by limiting sash travel to 20 Vertical rising sash may be raised from a closed to 20 operating height Exhaust volume and blower sizing is based on the 20 height...

Страница 23: ...e field mounted on the right corner post Receptacles are field wired to the single point junction box 13 Shipped fully assembled and eliminates the need for costly onsite assembly 14 Accessory Guardia...

Страница 24: ...Chapter 4 Performance Features Safety Precautions Product Service 1 800 522 7658 20 Original instructions Figure 4 1 16 18 7 17 5 15 14 11 10 9 7 12 8 2 1...

Страница 25: ...BURNED OUT LIGHT BULBS TO MAINTAIN MAXIMUM ILLUMINATION DO NOT OVERLOAD THE WORK SURFACE WITH APPARATUS OR WORK MATERIAL THE SAFE OPERATION OF THE LABORATORY HOOD IS BASED UPON HAVING PROPER AIRFLOW T...

Страница 26: ...the explosive limit are not recommended Mod les antid flagrants disposent composants lectriques antid flagrants Par cons quent l utilisation de mat riaux inflammables ou explosives dans des quantit s...

Страница 27: ...aque fois qu il est d plac l entretien ou au moins chaque ann e par la suite Ensure that the unit is connected to electrical service in accordance with local and national electrical codes Failure to d...

Страница 28: ...The surfaces of the motor can become hot and could cause burns Ne pas toucher le moteur de soufflerie Les surfaces du moteur peuvent devenir chaudes et pourrait causer des br lures Your Next Step Afte...

Страница 29: ...ur Fiberglass 30 Fume Hood utilizes a remote style blower or integral blower which can be activated by turning the blower switch to ON You can validate the hood performance by watching smoke drawn int...

Страница 30: ...ea Work Techniques Keep all materials at least 6 inches inside of the sash and perform all contaminated operations as far to the rear of the work area as possible Segregate all clean and contaminated...

Страница 31: ...you are not properly trained The wrench icon precedes the service operations that require qualified technicians Techniciens de certification Seulement form s et exp riment s devraient effectuer certa...

Страница 32: ...equipped for proper operation V rifiez que toutes les vannes de service si quip pour un bon fonctionnement The hood baffles should be checked for blockages behind them to ensure that the hood is main...

Страница 33: ...ash and reach inside the fume hood to unscrew the clear glass globe that protects the incandescent bulb 3 Install the new incandescent bulb and then reverse the removal procedure Do not contact blower...

Страница 34: ...rage Cabinets An optional work surface is available when installing the hood on storage cabinets Contact Labconco Customer Service for ordering information on these 30 work surfaces and storage cabine...

Страница 35: ...ray 2832608 Steam STM Black Installing Guardian Digital Airflow Monitor or Guardian Jr Airflow Monitor The Guardian Digital Airflow Monitor P N 9743201 and 9743205 or Guardian Jr Airflow Monitor P N 9...

Страница 36: ...proof hoods The following duplex kits are available Si vous avez command votre hotte et que vous voulez une sortie duplex lectrique alors vous pouvez avoir un install sur le terrain par un lectricien...

Страница 37: ...ck connection to control box on top of unit Circuit breakers tripped in building electrical supply Reset circuit breakers Remote blower won t operate but light works Blower wiring is disconnected Belt...

Страница 38: ...er exhaust system Hood should have average face velocity of 80 100 fpm Vertical rising sash no longer operates smoothly Cable is frayed or plastic protection is damaged Inspect cable and replace cable...

Страница 39: ...OWN 3F 1 9818805 Hose Barb Red Hot Water NOT SHOWN 3G 1 9818806 Hose Barb White Deionized Water NOT SHOWN 3H 1 9818807 Hose Barb Black Neutral or Steam NOT SHOWN 3I 1 9818808 Hose Barb Brown Nitrogen...

Страница 40: ...V EP 15E 1 1852300 Reducing Bushing 230V only 15F 4 1662600 Rubber Grommet 115V only 16A 1 1450000 Wheel Blower 115V only 16B 1 4716200 Wheel Blower 230V 115V EP 230V EP 17 1 2833400 Removable Side Pa...

Страница 41: ...Service 1 800 522 7658 37 Original instructions 24 12 32 20 28 Figure A 1 19 23 29 22 16 Not Shown 15 14 21 Not Shown 24 Not Shown 25 Not Shown 31 27 Not shown 26 9 10 Not shown 7 1 through 6 Not Show...

Страница 42: ...Product Service 1 800 522 7658 38 A AP PP PE EN ND DI IX X B B F FI IB BE ER RG GL LA AS SS S 3 30 0 H HO OO OD D D DI IM ME EN NS SI IO ON NS S Figure B 1...

Страница 43: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 39 Original instructions Figure B 2...

Страница 44: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 40 Original instructions Figure B 3 DIMENSION IN INCHES...

Страница 45: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 41 Original instructions Figure B 4 DIMENSION IN INCHES...

Страница 46: ...Appendix B Fiberglass 30 Hood Dimensions Product Service 1 800 522 7658 42 Original instructions Figure B 5...

Страница 47: ...e Transient over voltages according to Installation Categories II Over voltage Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 25...

Страница 48: ...Appendix C Fiberglass 30 Specifications Product Service 1 800 522 7658 44 Original instructions...

Страница 49: ...Appendix C Fiberglass 30 Specifications Product Service 1 800 522 7658 45 Original instructions...

Страница 50: ...ation ACGIH Fume hood face velocities between 60 100 lfpm Maximum of 125 lfpm for radioisotope hoods Duct velocities of 1000 2000 fpm for vapors gasses and smoke Stack discharge height 1 3 2 0 x build...

Страница 51: ...t air from fume hoods should not be recirculated Services should be external to the hood Canopy hoods only for non hazardous applications Materials of construction should have flame spread of 25 or le...

Страница 52: ...dbook of Laboratory Safety 2nd Edition CRC Press 1971 Department of Labor Occupational Safety and Health Administration 29 CFR Part 1910 Occupational Exposures to Hazardous Chemicals in Laboratories F...

Отзывы: