Chapter 2: Installation
Product Service 1-800-522-7658
9
Figure 4
Drain Connection
Using the supplied molded drain hose, connect and clamp the hose to the drain
outlet located at the bottom of the drain pipe. The drain hose should be routed to
an appropriate drain system, rated for a minimum capacity of 2 gallons (7.6 liters)
per minute. The RapidStill II’s drain pipe is designed for gravity drainage. The
drain hose and all connections must be below the level of the drain pipe outlet.
The drain hose connection to the drain system must meet all applicable local
standards for drain connections.
DO NOT restrict the drain hose or its
connections. Restrictions of the drain line may produce a pressure buildup
in the RapidStill II, resulting in damage to the unit and water leakage.
NaOH Pump Connection
Using either the silicone tubing supplied in the installation kit, or another
acceptable type of tubing, connect the NaOH pump inlet quick connector to the
tubing. Connect the connector to the pump inlet by pressing it straight in until the
connector ‘clicks’ and locks in place. The connector can be removed by
depressing the metal button on the body of the connector and pulling the
connector straight out. Place the other end of the tubing into the user-supplied
caustic container. Provisions should be made for the easy removal of the tube
from the caustic container, and its placement into a container of purified water for
rinsing of the system at the end of each work session.
Содержание 6520000
Страница 1: ...RapidStill II Models 6520000...
Страница 8: ...Chapter 1 Introduction Product Service 1 800 522 7658 5 Figure 2...
Страница 10: ...Chapter 1 Introduction Product Service 1 800 522 7658 7 Figure 3...
Страница 14: ...Chapter 2 Installation Product Service 1 800 522 7658 11 Rear View of Standard RapidStill II Figure 5...
Страница 29: ...Product Service 1 800 522 7658 26 A Ap pp pe en nd di ix x A A D Di im me en ns si io on ns s Figure 10...