Chapter 3: Normal Operation
Product Service 1-800-522-7658
14
(8)
When the distillation is completed, the unit will return to a
standby position, with the boiler heater still in operation.
(9)
Remove the digestion tube, using appropriate heat and chemically
resistant gloves.
(10)
If there are to be additional distillations and the boiler water level
is low, turn the ‘Boiler Heater’ switch off and depress the ‘Boiler
Water’ switch on. Allow the boiler to refill with water, release the
‘Boiler Water’ switch and turn the ‘Boiler Heater’ switch on.
When the water in the boiler resumes boiling, the RapidStill II is
ready for another distillation.
(11)
Titrate the distillate in the receiving flask. If the temperature of
the solution exceeds 120°F (50°C), check the flow rate of the
condenser cooling water.
Safety Precautions
The disposal of substances used in connection with this equipment may be
governed by various federal, state or local regulations. All users of this
equipment are urged to become familiar with any regulations that apply in the
user’s area concerning the dumping of waste materials in or upon water, land
or air and to comply with such regulations.
The RapidStill II should not be operated in damp conditions or those suspect
of an explosion and should be situated to allow adequate room and
ventilation during its normal usage.
The RapidStill II employs the following: electrical devices, heating element,
hot water, steam, pressurized water and steam lines. Care should be taken
when installing, operating, or servicing the unit.
Ensure that the RapidStill II is connected to electrical service in accordance
to local and national electrical codes. Failure to do so may create a fire or
electrical hazard.
The RapidStill II is designed to be connected to a cold, potable water supply
purified by municipal water treatment facilities. The installation of the
optional water inlet kit allows for the use of purified water to supply the
boiler. Never connect the RapidStill II to non-potable, unprocessed surface
or well water systems. The use of inadequate water supplies may damage the
unit.
Always wear appropriate eye protection, protective gloves and clothing when
using the RapidStill II.
Содержание 6520000
Страница 1: ...RapidStill II Models 6520000...
Страница 8: ...Chapter 1 Introduction Product Service 1 800 522 7658 5 Figure 2...
Страница 10: ...Chapter 1 Introduction Product Service 1 800 522 7658 7 Figure 3...
Страница 14: ...Chapter 2 Installation Product Service 1 800 522 7658 11 Rear View of Standard RapidStill II Figure 5...
Страница 29: ...Product Service 1 800 522 7658 26 A Ap pp pe en nd di ix x A A D Di im me en ns si io on ns s Figure 10...