5
3.
Operating Exclusions
Never use this instrument for
Liquids attacking FEP, PFA, PTFE or Hastelloy.
Liquids attacking borosilicate glass (e.g., hydrofluoric acid)
Liquids which are decomposed catalytically by Hastelloy (e.g., H
2
O
2
)
Explosive liquids.
Suspensions as solid particles may clog or damage the instrument.
4.
Operating Limitations
Concentrated hydrochloric acid and concentrated nitric acid, chlorinated and fluorinated
hydrocarbons or liquids which form deposits may make the piston difficult to move or may
cause jamming. When dispensing flammable media, make provisions to avoid static charging,
e. g., do not dispense into plastic vessels; do not wipe instruments with a dry cloth.
Warning!
If there is a sign of a potential malfunction (e.g., piston difficult to move) never use force.
Immediately stop dispensing and follow cleaning instructions or contact the manufacturer.
Note:
Compatibility of the instrument for this special application (e. g., trace material analysis) must
be checked by the user or contact the manufacturer.