background image

 

3.

 

Operating Exclusions 
 

Never use this instrument for 

 

Liquids attacking FEP, PFA, PTFE or Hastelloy. 

 

Liquids attacking borosilicate glass (e.g., hydrofluoric acid) 

 

Liquids which are decomposed catalytically by Hastelloy  (e.g., H

2

O

2

 

Explosive liquids. 

 

Suspensions as solid particles may clog or damage the instrument. 

 

4.

 

Operating Limitations 

 
Concentrated hydrochloric acid and concentrated nitric acid, chlorinated and fluorinated 
hydrocarbons or liquids which form deposits may make the piston difficult to move or may 
cause jamming. When dispensing flammable media, make provisions to avoid static charging, 
e. g., do not dispense into plastic vessels; do not wipe instruments with a dry cloth. 
 

Warning! 

If there is a sign of a potential malfunction (e.g., piston difficult to move) never use force. 
Immediately stop dispensing and follow cleaning instructions or contact the manufacturer. 
 

Note: 

Compatibility of the instrument for this special application (e. g., trace material analysis) must 
be checked by the user or contact the manufacturer. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 550.001.305

Страница 1: ...5 1 to 5mL 550 001 310 2 to 10mL 550 001 325 2 5 to 25mL 550 001 350 10 to 50ml Please read the User Manual carefully before use and follow all operating and safety instructions Technical specificatio...

Страница 2: ...onents 4 3 Operating Exclusions 5 4 Operating Limitations 5 5 Preparing instrument for use 6 6 Mounting instrument on bottle 8 7 Priming the instrument 8 8 Cleaning 11 9 Autoclaving 12 10 Checking the...

Страница 3: ...lth practices and determine the applicability of regulatory limitations prior to use a Follow the general instructions for hazard prevention and safety regulations e g wear protective clothing eye pro...

Страница 4: ...cap 16 Valve block 1 2 3 4 7 8 9 11 10 12 13 14 5 6 15 16 Application and Design This instrument is designed for dispensing liquids observing the following limits 15 to 40 C of instrument and reagent...

Страница 5: ...luorinated hydrocarbons or liquids which form deposits may make the piston difficult to move or may cause jamming When dispensing flammable media make provisions to avoid static charging e g do not di...

Страница 6: ...h does not attack PP Warning Wear protective clothing eye protection and gloves Follow all Safety Instructions and observe the Operating Exclusions and Limitations Figure 1 Warning Always wear protect...

Страница 7: ...charge tubes 1 2 Figure 2 1 Firmly tighten the discharge valve 2 Push the discharge tube as far as possible into the discharge valve Connecting the filling tube 1 2 Figure 3 1 Cut off the bottom of th...

Страница 8: ...e made of polypropylene PP and can only be applied to media which do not attack PP 7 Priming the instrument Attention For small bottles use a bottle stand to prevent tipping over Warning Reagent may d...

Страница 9: ...e piston up then press it down rapidly 3 Repeat this procedure until most of the air bubbles in the glass cylinder and discharge tube have been removed Note Before using the instrument for the first t...

Страница 10: ...s pushed on Reagent may drop off the discharge tube and closure cap Setting the volume 1 2 3 Figure 7 Figure 8 Hold an appropriate collecting vessel under the discharge tube opening Pull the piston up...

Страница 11: ...of the liquid but still in the bottle 3 Carefully tap the filling tube against the bottle from the inside so that the reagent runs back into the bottle 4 Remove the closure cap and dispense the remain...

Страница 12: ...2 Unscrew the discharge tube 3 Remove valve locking ring 4 Unscrew the discharge valve using the mounting tool Remove the valve with its sealing washer Make sure the sealing washer does not remain in...

Страница 13: ...autoclaving inspect all parts for deformities or damage If necessary replace them Do not reassemble the instrument until it has cooled down to room temperature cooling time approx 2 hours Mounting to...

Страница 14: ...mble Follow all cleaning instructions Filling not possible Volume adjusted to minimum setting Sticking filling valve Set to required volume Clean the filling valve If necessary replace the valve with...

Страница 15: ...original manufacturer s tube on firmly Cleaning procedure Push the filling tube on firmly If this does not suffice cut off approx 1 cm of tube at the upper end and re connect it or replace filling tub...

Страница 16: ...16...

Отзывы: