Firmware
version
Firmware description
1.00.06
dd. 15.01.07.
Final firmware
E20-10
rev.
А
.
2.00.03
dd. 10.04.08.
First firmware of serial
E20-10
rev.В.
Users applying internal calibration in
E20-10
rev.В are recommended to update firmware
to elder version where the effect to damage individual data samples in code approximating
to ADC word size edge has been eliminated. Negative effects have not been found when
internal calibration is off.
2.00.05
dd. 24.07.08.
Updated firmware of serial
E20-10
rev.В.
In FPGA 2.00.05 firmware the additional program start is not required under
additional startup condition "on
signal passing through given level in selected channel"
and shutdown condition "on number of M recorded frames", due completing of the next
shutdown condition fulfillment the additional data acquisition start will begin
automatically upon completing of new start condition fulfillment. For another start
conditions
FPGA 2.00.05 firmware is similar to 2.00.03 firmware. Also refer to
FPGA 2.00.03 and 2.00.05 firmwares are fully compatible on program library functions.
2.00.06
dd. 01.09.08г.
Updated firmware of serial
E20-10
rev.В.
One-channel mode sensitive data acquisition fault has been eliminated
.
1.00.07
dd. 29.10.08.
Updated firmware of serial
E20-10
rev.A.
Data corruption in repetitive interleaving of external start and program shutdown has been
eliminated (this relates to rev.A only). FPGA 1.00.07 firmware means simultaneous
upgrading of AVR controller firmware.
2.00.07
dd. 05.09.12.
For E20-
10 rev.В the
embedded software upgrading enabling to boost
maximum possible
frame size up to 8192 samples of
ADC
has been developed! To upgrade it is required to
alter AVR controller and upgrade library remotely (FPGA 2.00.07 firmware has been yet
added in which). Compatibility with old software of upper level is existed.
2.00.08
dd. 20.11.12.
Firmware for E20-
10 rev.В. In comparison with version 2.00.07, the fault on odd data
interference in ADC buffer in additional data acquisition starting on signal START under
data acquisition program shutdown mode has been eliminated.
2.00.10
dd. 15.12.14.
Firmware for E20-
10 rev.В. Data acquisition restart on signal START fault has been
eliminated.