6.1.8. The coordinated connection of remote sources of current or voltage through a
long line with a wave impedance Zw
(If the signal source has an output impedance, unequal to Zw, then these cases correspond to one-way
matching of the long line on the side of the signal receiver)
6.1.8.1. Mode "with common ground"
6.1.8.2. Differential mode
6.1.8.3. Differential connection of an isolated current source or voltage
R1=R2
R1+R2 = Zw
The circuit is valid only for the following E-502
voltage sub-ranges:
±2 V, ±1 V, ±0.5 V, ±0.2 V
R=Z
в
E-502
AGND
GND32
X1 (X2-X16, Y1-Y16)
Z
в
Y1 (Y2-Y16)
E-502
AGND
X1 (Y2 -Y16)
Z
в
R=Z
в
R1
E-502
AGND
X1 (X2 -X16)
Y1 (Y2 -Y16)
R2
Z
в