6.1.5. Differential connection of the transformer (throttle) winding with midpoint and
offset potential with respect to AGND
Limitations:
|Ucm| ≤ 1 V
Only for E-502 voltage subbands:
±2 V, ±1 V, ±0.5 V, ±0.2 V
6.1.6. Example of mixed connection of voltage sources "with common ground" and
differential.
6.1.7. Connecting a power supply to the ADC input
6.1.7.1. Mode "with common ground"
6.1.7.2. Differential mode
Fixed ADC ± U subband must correspond to U=I
MAX
*R, and the current source must have a voltage margin
of at least U. Resistor R should always be located close to the ADC input. In any case, the resistor R must
be less than 50 Ω in the multi
-channel mode at the maximum switching frequency.
R1
E-502
AGND
X1 (X2 -X16)
Y1 (Y2 -Y16)
L
U
см
GND
E-502
GND32
X (Y )
i i
GND1
GND3
Yk
X k
X (Y )
j j
Yr
X r
GND2
Дифф.
ИН
R
I
E-502
AGND
GND32
X1 (X2-X16, Y1-Y16)
Y1 (X2-X16)
R
I
E-502
AGND
X1 (Y2 -Y16)