background image

 

 

CAUTION 

●CAUTION 

Be sure to observe the caution and precautions described in the EDX-10B Instruction Manual (FOR 

HARDWARE). 

 Do not use the EDX-10B outdoors. 

Or, it may cause electric shock, fire hazard, lower the performance and cause trouble. 

 Use the EDX-10B within temperature ranging from 0 to 40ºC. 

Use at temperatures exceeding the specified range may lower the performance and cause trouble. 

If use under direct sunlight or in a cold place is inevitable, prepare a sunscreen or take proper measures to keep 

it warm. 

 Use the EDX-10B in the specified operating humidity of 20 to 85% RH. 

Use in a humid place exceeding the specified range or where it is exposed to splashing water may lower the 

performance and cause trouble. 

 Do not use the EDX-10B immediately after the change in the environment. 

Leave the EDX-10B as it is to acclimate to the environment. 

Abrupt change in ambient temperature due to transportation, etc. may cause dew condensation, which may 

result in lower performance and troubles. 

 Do not use the EDX-10B where it is subject to significant vibration or impact. 

The vibration resistance and impact resistance are as follows. 

Vibration resistance:   

29.42m/s

2

 (3G), 5 to 200Hz (when operating) 

Impact Resistance:   

294m/s

2

(30G)/11msec 

Use the EDX-10B in an area where vibration and impact can be kept within the scope of specifications. 

Continuous vibration or severe impact may cause deteriorated performance and system failure. 

 Do not use the EDX-10B in strong electromagnetic field. 

Use the EDX-10B in an environment where magnetic field is acceptable to the PC. 

Performance may be lowered and erroneous operation and troubles may result if it is used near a telemetry 

system, microwave oven, electronic furnace or any other equipment generating a strong magnetic field. 

 Do not use the EDX-10B under poor conditions. 

Optional AC adapter (UN310-0515): 100 to 240 VAC (50 Hz/60 Hz) The power supply used for the EDX-10B 

should not experience an instantaneous power failure and be free from noise. 

 The specifications of the optional AC adapter (UN310-0515) is as follows. 

Input voltage:     

 

 

100 to 240 VAC (50 Hz/60 Hz) 

Output voltage:   

 

 

5 VDC (1.5 A) 

Operating Temperature Range 

 

0 to 40

 

 The optional AC adapter (UN310-0515) is consumable item. Replace it every five years. 

 Do not connect the USB cable immediately after disconnecting. 

Wait approximately 5 seconds before connecting the USB cable again. 

 Do not pull cables. 

Lay cords and cables with a certain allowance so that unreasonable force is applied to the connections. 

Pulling or unreasonable force may cause troubles or interrupt the measurement. 

 Avoid installing sensors and EDX-10B near a welding machine. 

Failure to do so will pose the risk of erroneous data, malfunction and failure. 

 Do not disassemble or remodel the EDX-10B. 

Disassembling or remodeling by user may cause electric shock hazards or damage the EDX-10B. 

This warranty does not cover any damaged or defective parts that results from disassembling or remodeling. 

 Do not use the EDX-10B under dusty environment. 

This can lead to performance problems and decreased operating efficiency. 

Be careful that dust may not enter the EDX-10B not only during operation but also under stored conditions. 

 Preheat the EDX-10B before use. 

After the power ON, always preheat the EDX-10B for approximately 10 minutes. 

Содержание EDX-10B

Страница 1: ...the DCS 100A Instruction Manual For EDX 10A B Operation Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and or other countries Generally company names and trade names described in this Instruction Manual are trademarks or registered trademarks of the companies This Instruction Manual may not be copied or reproduced in whole or part...

Страница 2: ......

Страница 3: ... AND INDICATORS 9 2 1 FRONT 9 2 2 SIDE 9 2 3 BOTTOM 10 3 CONNECTION AND DEFAULT SETTING 11 3 1 PROCEDURES FOR MEASUREMENTS 11 3 2 CONNECT THE MEASURING UNITS 12 3 2 1 NUMBER OF CONNECTABLE MEASURING UNITS 12 3 2 2 HOW TO CONNECT THE MEASURING UNIT 13 3 3 CONNECT SENSORS 13 3 4 CONNECT THE OPTIONALAC ADAPTER 14 3 5 INSTALLING THE USB DEVICE DRIVER 15 3 6 CONNECTING THE USB INTERFACE 16 3 6 1 Connec...

Страница 4: ...00A 1 CD R Ground wire P 72 1 USB cable N 38 1 Rubber Foot 8 Test Certificate and Warranty 1 each EDX 10B READ IN THE BEGGNING English Japanese 1 EDX 10B READ IN THE BEGGNING Chinese 1 OPTIONAL ACCESSORIES AC adapter UN310 0515 Strain measuring unit EDX 11A Voltage measuring unit EDX 12A Thermocouple measuring unit EDX 13A Low power strain measuring unit EDX 14A Low power strain voltage measuring ...

Страница 5: ...n the Instruction Manual WARNING Improper operation of the system may result in death or severe injury of the operator CAUTION Improper operation of the system may result in injury of the operator and physical damage of the system WARNING WARNING Be sure to observe the warning and precautions described in the EDX 10B Instruction Manual FOR HARDWARE Power Supply To prevent a fire hazard when using ...

Страница 6: ...eptable to the PC Performance may be lowered and erroneous operation and troubles may result if it is used near a telemetry system microwave oven electronic furnace or any other equipment generating a strong magnetic field Do not use the EDX 10B under poor conditions Optional AC adapter UN310 0515 100 to 240 VAC 50 Hz 60 Hz The power supply used for the EDX 10B should not experience an instantaneo...

Страница 7: ...ity Do not apply pressure to the CD R by laying object or bending Handle with care to avoid any fingerprint dirt scratch on both sides Cleaning When the product gets dirty clean the product with a dry soft cloth When dust exists inside the EDX 10B clean it by using an air blow gun Do not touch electronic parts ...

Страница 8: ...l be emitted resulting in an unstable voltage To operate the EDX 10B normally finding a sign of the EDX 10B failure early by daily periodic inspections and taking the corrective action are required For inspection contact KYOWA or our representatives All components do age and will fail EDX 10B REPLACEMENT BEFORE THE END OF SERVICE LIFE Preventive maintenance and replacement parts are a cost effecti...

Страница 9: ...nnect 5 or more measuring units Or deterioration or malfunction of the instrument may result 5 Be sure to use the accessory screws binding head screw M3 5 for coupling measuring units Longer screws will damage the inner PCB components 6 Be sure to use standard accessory USB cable Or deterioration or malfunction of the instrument may result 1 Be sure to use standard accessory USB cable 2 Be sure to...

Страница 10: ...not used EDX 11A Applicable for high speed measurement with the maximum sampling frequency of 20k Hz up to 4 channels When connecting 4 measuring units 16 channels the maximum sampling frequency is 5k Hz Connection with the PC is easy with the use of the USB interface You can measure and record data easily by using the Dynamic Data Recording Software DCS 100A The Dynamic Data Acquisition Software ...

Страница 11: ...he LED lights up in red stop using the EDX 10B and disconnect the USB cable after disconnecting the optional AC adapter When the number of connected units exceeds the number of connectable units The number of connectable units is refer to 3 2 1 NUMBER OF CONNECTABLE MEASURING UNITS When connecting 5 or more measuring units Voltage range of the optional AC adapter exceeds 4 5 to 5 5 V Voltage range...

Страница 12: ...ubber feet optional to keep it from sliding around when placed on a flat surface 9 Cap Cap for protecting the unit coupling connector on the EDX 10B bottom Remove the cap before connecting measuring units 10 Unit coupling Connector Connector for connecting measuring units 8 10 9 ...

Страница 13: ... to the EDX 10B Section 3 3 Connect sensors Connect various sensors to the measuring units Section 3 4 Connect the optional AC adapter Connect the optional AC adapter Ground the GND terminal Section 3 5 Installing the USB device driver Install the USB device driver into the PC Section 3 6 Connect the USB interface Set the USB interface for connecting the PC Chapter 4 Measure data Preparation and m...

Страница 14: ...EDX 11A the number of connectable measuring units is 4 units When used EDX 11A If you don t use EDX 11A When used USB bus power USB specification 2 0 When used USB2 0 bus power the number of connectable measuring units is change by type of measuring units When used EDX 11A the number of connectable measuring units is 1 unit only one EDX 11A If you don t use EDX 11A the number of connectable measur...

Страница 15: ...ng unit for connction Or the holding pin and connectors on the measuring unit will be damaged 3 2 2 HOW TO CONNECT THE MEASURING UNIT This section describes how to connect the measuring unit into the EDX 10B 1 Remove the cap on the EDX 10B bottom 2 Connect the unit coupling connector on the EDX 10B bottom and connector on the measuring unit top Use the finger slots to connect the EDX 10B and a mea...

Страница 16: ...connected four measuring units If you want to use USB bus power plese go to 3 5 INSTALLING THE USB DEVICE DRIVER 1 Connect the DC plug of the optional AC adapter into the DC IN connector of the EDX 10B 2 Connect the optional AC adapter into the AC power outlet The power supply voltage should be checked according to the procedure described below 1 Use the tester to measure the voltage between the A...

Страница 17: ...o the EDX 10B by USB cable 2 Please log in as administrator 3 Insert the CD into the PC 4 Double Click the UsbDriver KYOWA_USB_ exe in the CD is version of driver 5 Follow the instruction on screen Finish button appears when the setup completes 6 Please connect the PC to EDX 10B by USB cable Installation of the USB driver will start For details see Instruction Manual FOR USB DRIVER SET UP which is...

Страница 18: ...10B cannot measuring Do not connect other machine to USB port Or deterioration or malfunction of the instrument may result 3 6 CONNECTING THE USB INTERFACE 3 6 1 Connect the USB cable 1 Connect the B connector of the USB cable into the USB connector of the EDX 10B 2 Connect the A connector into the USB port of the PC 3 Make sure the LED lights up in green after starting up the Dynamic Data Recordi...

Страница 19: ...ssory USB cable Or deterioration or malfunction of the instrument may result 3 6 2 Wiring diagram of the USB interface 1 When connecting the PC and EDX 10B directly 2 When connecting the PC and EDX 10B via USB hub Be sure to use AC adapter when using USB hub Compliant to the USB specification 2 0 A connector B connector A connector Mini B connector Use the high speed USB hub compliant to the USB s...

Страница 20: ...er you cannot use USB hub Or deterioration or malfunction of the instrument may result 3 When connecting the PC and EDX 10B directly Use USB bus power When you use USB bus power USB specification 3 0 When you use USB bus power USB specification 3 0 Use the high speed USB hub compliant to the USB specification 3 0 A connector B connector USB cable Use the high speed USB hub compliant to the USB spe...

Страница 21: ...or EDX 10A B Operation 5 Monitor the data on the PC for checking the input signal Conduct the balancing operation if required If noisy set the low pass filter on the channel condition setting window and check the grounding 6 Start recording data Start recording data by using the Dynamic Data Recording Software DCS 100A The recorded data is saved in the specified folder in the PC 7 Stop recording d...

Страница 22: ...ermeasures contact KYOWA or our representative If the EDX 10B is damaged due to disassembling or remodeling by the user KYOWA may refuse the repair Trouble Checked Items and Countermeasures The LED blink in red after starting up the Dynamic Data Recording Software DCS 100A The number of connected units exceeds the number of connectable units The number of connectable units refer to 3 2 1 NUMBER OF...

Страница 23: ... life parts Item Content Service intervals AC adapter OPTIONAL ACCESSORIES The optional AC adapter UN310 0515 is a limited part Service life depends on the frequency of use and ambient temperature is approximately 5 years We recommend you purchase a new AC adapter every 5 years Once in every 5 years The replacement frequency of limited life parts is about five years assuming eight hours of use 365...

Страница 24: ... Approx 170 g 10 Dimensions 84 0 W 26 6 H 84 0 D mm excluding protrusions 11 Control Software DCS 100A 12 Applicable Standard EMC Directive EN61326 1 Class A RoHS Directive EN50581 1 When using USB bus power please connect directly to the USB port of the PC without using a USB hub 2 The number of measurement units at USB bus power operation refer to the following table USB Port The number of EDX 1...

Страница 25: ...23 8 OUTSIDE DRAWING 8 1 EDX 10B Unit mm ...

Страница 26: ...24 8 2 WHEN CONNECTING FOUR EDX 11A UNITS Unit mm ...

Страница 27: ...25 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...

Отзывы: