19
4. MEASURE DATA
4-1. PROCEDURE
(1) Install the Dynamic Data Recording Software DCS-100A (for the first time only).
Install the Dynamic Data Recording Software DCS-100A into the PC.
For details, see the DCS-100A Instruction Manual (For Common Operation).
(2) Mount the EDX-10B.
Mount the EDX-10B.
For details, see 3.CONNECTION AND DEFAULT SETTING.
(3) Turn ON the PC.
Turn ON the PC to start up Windows.
Connect the PC and EDX-10B by referring to 3-6. CONNECTING THE USB INTERFACE.
(4) Set the measuring conditions.
Start up the Dynamic Data Recording Software DCS-100A. Set the channel conditions, measuring conditions, and
sampling frequency.
The LED will light up in green after starting up the Dynamic Data Recording Software DCS-100A.
For details, see the DCS-100A Instruction Manual (For EDX-10A/B Operation).
(5) Monitor the data on the PC for checking the input signal.
Conduct the balancing operation if required.
If noisy, set the low-pass filter on the channel condition setting window and check the grounding.
(6) Start recording data.
Start recording data by using the Dynamic Data Recording Software DCS-100A.
The recorded data is saved in the specified folder in the PC.
(7) Stop recording data
Stop recording data by using the Dynamic Data Recording Software DCS-100A.
(8) Check the recorded data waveform
Check the recorded data waveforms by using the Dynamic Data Recording Software DCS-100A.
After retrieving data, you can check and analyze the recorded data by using the Dynamic Data Acquisition Software
DAS-200A.
4-2. EDX-10B POWER ON/OFF
The EDX-10B power ON when the USB cable is connected and power OFF when the USB cable is disconnected.
To turn OFF the EDX-10B after measurements, disconnect the USB cable after disconnecting the optional AC adapter.