Fully Automated FOB analyzer: HM-JACK 

series

 

 

 

 

 

 

 

 

 

FECAL OCCULT BLOOD TEST ANALYZER 

For Professional Use only 

 

Operation Manual

Rev.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Kyowa Medex Co., Ltd. 

 

 

IVD 

Содержание HM-JACK Series

Страница 1: ...l i Fully Automated FOB analyzer HM JACK series FECAL OCCULT BLOOD TEST ANALYZER For Professional Use only Operation Manual Rev 4 Kyowa Medex Co Ltd IVD ...

Страница 2: ...ed to Main Program Ver 3 01 Rev 3 Feb 2016 6 3 Added some description to List of the Last three Digits of Error Codes Rev 4 Oct 2017 7 1 Added an item to specifications of Power supply Fully Automatic Human Fecal Occult Blood Hemoglobin Analyzer HM JACKarc Operation Manual First edition published December 2011 Written and edited by Medical Instrument Department Kyowa Medex Co Ltd Published by Kyow...

Страница 3: ...ng Symbol 2 1 2 2 Explanation of Warnings 2 1 2 3 Explanation of Warning Labels Affixed to the Analyzer 2 2 2 4 Safety Precautions for Operators 2 3 2 5 Safety Precautions for Maintenance 2 5 2 6 Waste Materials Precautions 2 6 2 7 Instrument Safety Device 2 7 2 8 Noise Level of the Analyzer 2 7 2 9 Note for Supplied Water and Detergent 2 7 3 Outline of Analyzer 3 1 3 1 Characteristics of the Anal...

Страница 4: ...matically from Sleep Mode 5 2 Logging in to the Analyzer 5 3 5 2 2 Preparing Detergent in the Detergent Tank 5 4 5 2 3 Preparing the Waste Tank 5 5 5 2 4 Preparing the Reagent 5 6 Setting the Reagent in place 5 6 Setting Reagent volume 5 7 5 2 5 Preparing the Buffer 5 9 5 2 6 Preparing the Reaction Cell 5 12 5 2 7 Registration of Master Curve and Calibration Card 5 14 5 2 8 Preparation of Samples ...

Страница 5: ...formation of Master Curve 5 28 Inputting Lot information of Master Curve 5 28 5 5 3 Setting the first position of picking up the Cell 5 30 5 5 4 Setting Printer Paper 5 31 5 5 5 Actions for Recovery from Emergency STOP 5 33 6 Error Messages 6 1 6 1 Structure of the Codes 6 1 6 2 List of the First Two Digits 6 2 6 3 List of the Last Three Digits of Error Codes 6 3 6 4 Trouble Shooting Actions for t...

Страница 6: ...ontent of the Manual and software is subject to change without notice Warranty period Refer to local distributor Warranty coverage During the warranty period all malfunctions due to manufacturing or design faults will be repaired free of charge Exemptions The following are not covered by warranty even if they occur within the warranty period 1 Malfunctions due to disassembly or modifications not c...

Страница 7: ...ject to change without notice While we have taken all possible precautions to ensure quality in the content of this Manual please contact your dealership or contact Kyowa Medex Co Ltd if you find anything unclear or any apparent errors or omissions Our contact details are stated on the last page of this Manual This Manual is copyrighted by Kyowa Medex Co Ltd 2005 Kyowa Medex Co Ltd All Rights Rese...

Страница 8: ...han this Manual read the following operation manuals thoroughly and be sure to understand the information before performing analysis operations or handling the reagent and other materials Keep the Manual near the Analyzer If you are unsure of the analysis operation method for this Analyzer read this Manual thoroughly before going on with operation and analysis Store this Manual carefully Operation...

Страница 9: ...ng Precautions and handle the Analyzer accordingly This Manual indicates information that the user should be particularly aware of when performing analysis operation or maintenance management tasks as Request Indicates section to reference in this Manual Provides supplementary information 1 Numbers in italics in the text indicate the step numbers in procedures and the sequences of operations Perfo...

Страница 10: ...k of adverse influence from atmospheric pressure temperature humidity drafts sunlight dust air containing substances such as salinity sulfur etc or other environmental factors 3 Make sure the Instrument is safe from tilting vibration and impact including that during movement 4 Do not install the Instrument in a location where chemical substances are stored or where gas is generated 5 Pay close att...

Страница 11: ...perature humidity drafts sunlight dust air containing salinity and or sulfur or other environmental factors iii Make sure the Instrument is safe from tilting vibration and impact including that during movement iV Do not store the Instrument in a location where chemical substances are stored or where gas is generated 4 Clean the accessories cables and probes 5 Be sure to keep the Instrument clean 5...

Страница 12: ... and in this Manual These warnings are used separately as described below to indicate the degree magnitude of personal injury and property damage which could occur This signal and the legend is used in situations in which incorrect handling could result in serious injury to the user This signal and the legend is used in situations in which incorrect handling could result in minor injury to the use...

Страница 13: ...ng Labels Affixed to the Analyzer Make sure that warning labels are legible at all times If a label peels off contact local distributor immediately Never peel warning labels off or modify them The warning labels shown below are affixed to warn operators of potential hazards Caution ...

Страница 14: ...ble is not caught by pressure between the Instrument and the wall the cable is not bent too tight nothing is put on the cable the cable is not tied in a bundle the cable is not modified the cable is kept away from any heater the plug itself not the cable is gripped when unplugging the cable Dust accumulated on the power plug can start a fire Check the plug regularly and remove dust Insert the powe...

Страница 15: ...socket Leaving faults unattended can cause accidents and fires Do not turn the power switch off and on repeatedly The Mixing Spatula Sample Pipette and Reagent Pipette attached to the injection assembly are sharp and could cause injury The ends of the Mixing Spatula Sample Pipette and Reagent Pipette attached to the injection assembly come into direct contact with Samples stool samples and Reagent...

Страница 16: ...es could cause the Instrument to malfunction or in some cases start a fire Using the power cable plug plugging it in and unplugging it to turn the power on and off may cause the plug to overheat Always use the power switch or the Main Breaker This Instrument weighs approximately 56 kg Handle it with extra care when moving Note that moving the Instrument should be done by two or more people The end...

Страница 17: ... samples stool samples and reagents Incorrect handling of those parts could cause infection with pathogens or poisoning with reagents Wear medical rubber gloves at all times while working and take extra care to avoid infection Here below is a list of Waste materials generated by regular analysis They should be handled as medical waste and disposed of by a specialist contractor Liquid waste Used ce...

Страница 18: ...ot start motion If the outer cover is opened while the Analyzer is in motion an error occurs and the Analyzer will stop sampling There are 2 switches to secure the function 2 8 Noise Level of the Analyzer The maximum noise level is 70db 2 9 Note for Supplied Water and Detergent Temperature of supplied water The temperature of supplied water for this Instrument should be in 20 30 Lower temperature ...

Страница 19: ...t calibration method is adopted to simplify the measurement operation The measurement method is Latex Agglutination which is quick and achieves superior sensitivity Superior detection sensitivity realizes smaller volume of stool sample and less smell The sample loading capacity of the Analyzer is 80 Reaction Cells are replaced after use automatically The Sample Collection Picker achieves superior ...

Страница 20: ...e to mix the Reagent and the Sample together The Cell containing the Reagent and the Sample moves to the detection position after a period of time set under the measurement conditions and the first photometric measurement ISTT1 is performed After another set period of time the second photometric measurement ISTT2 is performed The amount of change in the Integrating Sphere Turbidity of the two phot...

Страница 21: ...Operation and Display Panel backlight color LCD panel Printer Paper Microcomputer Interface Injector Buffer Fluid Latex Reagent Injection Arm Wash Station Turbidity Meter Cell Cassette Reaction Table Cell Transportation Supply and Waste Water Pump Waste Liquid Detergent Sample Transportation Sample Rack ...

Страница 22: ...he most of the characteristics of the integrating sphere photometer we have achieved Optimization of the volume of stool sample by the two step rotary scraping method Adoption of hexagonal stool sampling holes that improves both stool holding and solubility One way collection of excess stool and easy check of whether stool has been sampled Time saving by adopting integrated multifunction seals The...

Страница 23: ...s Instrument are as follows Top Right side Display Cell Cassette 40 Reaction Cellsx2 L R Outer Cover Sample Tray eject Sample Tray load Printer Injection Arm Handheld Scanner Reaction Table Reaction Cell Integrating Sphere Turbidity Meter Buffer Reagent Cover Latex Buffer bottle Inside the Cover USB Slot ...

Страница 24: ...l Outline of Instrument 3 6 Rev 4 Left side Back Cell Waste Box Cell Hand Cover Handheld Scanner Connector Cooling Fan Lamp Replacement Panel Breaker Switch Power Cable Plug Detergent sensor Detergent tube Waste tube Connectors ...

Страница 25: ...e list of required materials other than parts provided with the Analyzer EXTEL HEMO AUTO HS Latex EXTEL HEMO AUTO Buffer EXTEL HEMO AUTO HS Calibrator EXTEL HEMO AUTO HS Control EXTEL HEMO AUTO MC Collection Picker EXTEL HEMO AUTO Reaction Cell Please contact your local supplier to order the products ...

Страница 26: ...ial Screen This screen is displayed after the power is turned on Touch the screen to move to the Main Menu screen 4 1 2 Main Menu screen This screen is main screen to start operations Touching any RETURN key on each screen several times finally returns to this screen Applied Calibration curve Number Name of current screen Supplementary name of current screen Current versions of programs Main Contr...

Страница 27: ...con Touch this icon to return all moving parts to their home positions Calibration curve icon Touch this icon to go to Calibration screen Master Curve Calibrator registration are available from the screen See 4 1 5 Calibration for details Results icon Touch this icon to display analysis results The Results screen can be used to display or print results or send them to the host computer To display ...

Страница 28: ... System button From the menu various system parameters can be modified but it is not allowed to modify the parameters without permission Maintenance button Touch this button to go to the Maintenance Menu screen Refer to the Service Manual for details USB Mark if connected When a USB is inserted to the USB slot and recognized by the system the mark on the left appears Touching this icon allows the ...

Страница 29: ...n Sample Stop icon Touch this icon to stop sampling Ongoing measurement still goes on after sampling stop and will complete Home Position icon Touch this icon to return all moving parts to their home positions Sample Status field After starting measurement Sample Status will be displayed in this field See Sample Status field on page 4 7 ID column Not in Use Attribute field Not in Use Applied Calib...

Страница 30: ...easurement For actions after an emergency stop refer to page 5 28 Actions for Recovery from Emergency STOP Applied Calibration curve Number The currently applied Calibration curve Number No 1 in this case is displayed Name of current screen The name of the current screen is displayed Supplementary name of current screen The current operation status is displayed System button From the menu various ...

Страница 31: ...HM JACKarc Operation Manual Functions of Each Screen 4 6 Rev 4 Feed button Touch this icon to feed the printer paper H Copy Touch this icon to print out the hardcopy of the screen ...

Страница 32: ...he measurement again Main Menu icon Touch this icon to go back the Main Menu screen Rack Monitor Sampling The sample rack Number and position Number being sampled at the moment is displayed R ID of the Rack is displayed The position being sampled now is colored green Sampling End R ID of ejected Rack is displayed Status and operation message Applied calibration curve number Name of current screen ...

Страница 33: ...rameter 1 3 Control H Repetition number of Control H Set as an item parameter 1 3 Cup Repetition number of Sample Cup as optional Set as an item parameter 1 3 Patient Samples one day method 1 Patient Samples two day method 1 1st day 2 2nd day Data Results of the Sample is displayed Alarm If an event occurred for the Sample a flag is displayed in this column Refer to Error List for details of the c...

Страница 34: ...plementary name of current screen The current operation status is displayed System button Various parameters can be modified from this menu but it is allowed to enter this menu only when it is necessary to change input data for Calibrator manually Refer to for details To Order button Touch this button to start measurement after finishing previous one Start icon will appear USB Mark if connected Wh...

Страница 35: ...ome positions Position Status field The status of each position of the Reaction Table is displayed in five abbreviations as below OK A Cell is set and available MS Measurement is ongoing NG The position is checked but not available The position is not checked yet Blanc No Cell is present All Changes icon Touch this icon to replace all cells on the Reaction Table Selective Change icon Touch this ic...

Страница 36: ...en Cell Cassette screen Cell Cassette Map This map indicates presence of cells 1 40 on the cassettes R and L on the Analyzer White blocks A Cell is present Dark gray blocks No Cell is present Blue Block The position will be picked up first It can be changed with touching another one Start Position The position picked up first is displayed Cell Set icon Touch this icon to go back to Cell Set screen...

Страница 37: ...mpleted Lot information field Lot information of Master Curve and Calibrator is displayed Master button Touch this button to refer to the data input with registration of Barcodes or to input data manually Refer to for procedure for manual input of the data Curve button Touch this button to view the calibrated Master Curve Barcode button Touch this button to move to Barcode Registration mode See 5 ...

Страница 38: ... Rev 4 Curve 1 2 button Touch this button to change Calibration data to apply Confirmation window will appear after touching this button This is the only way to change Applied Calibration between 1 and 2 RETURN button Touch this button to go back to Main Menu ...

Страница 39: ...ct as the target mark will be displayed No Sample number Pos Rack ID and position number User ID The number that the system applies to every sample following a configuration in User Parameter Barcode Barcode of the Sample Touch this area to view Reaction Data See 4 1 6 Reaction Data for reference Av Number of repetition of Calibrator and Control Data Result of the Sample ng mL Qualitative Result M...

Страница 40: ...the selected Result s Calc button Touch this button to calculate results again When this button is touched two buttons will appear Recalc This button is effective only when the calibration data is revised The data can be calculated based on the new calibration data CV CV can be calculated from selected results Touch sel column of target data to select and touch Execute button The Calculated CV can...

Страница 41: ...is displayed Reaction Data field Reaction Data I S T of each point is displayed Reaction Curve field Reaction Curve of is displayed Lot Info button Reaction Data field will be changed to the information on the right PREV NEXT button Touch these buttons to view previous next result Measurement Information field Reaction Data field Reaction Curve field ...

Страница 42: ... control data and displayed The maximum number of calculated data is 31 from the start date The data plotted on the chart are average of each control result of a run Date button To direct the start date of the chart touch these buttons to display Numeric keys Enter the start date yyyy mm dd and touch Enter key Information field Statistic data are displayed Chart The Chart of each Low High control ...

Страница 43: ...ain data from the chart List of the data will be shown then touch Avail and mark with X to exclude from the High chart Histogram Result List When the Histogram button is touched the screen below is displayed Choose All Today or touch Sel to target for execution Sel column Target Data with mark All button Target All the data Today button Target Data of current date Graph button Tabulate the Graph i...

Страница 44: ...HM JACKarc Operation Manual Functions of Each Screen 4 19 Rev 4 4 2 System Menu Screen The System Menu is displayed when System button is touched These menus are for service personnel ...

Страница 45: ...u Also some configuration can be adjusted and the status of sensors can be checked This function is for service use DETECT TEST When this button is touched DETECT screen is displayed The scattering light and transmitted light can be observed on the screen This function is for service use Error History Error History display can be seen from this menu Maintenance Touch this button to go to the Maint...

Страница 46: ...rature before use Set the reagent volume on the Sample Order screen if necessary 5 Prepare the Buffer Let the Buffer warm to room temperature before use 6 Prepare the Reaction Cells If required use the Cell Set screen to set the starting position for picking up the cells and replacing used cells 7 Prepare for calibration When Latex lot is changed register Master Curve When the Calibrator lot is ch...

Страница 47: ...eating leading to fire and or electric shock It takes approximately one hour for the instrument to warm up and to be ready for analysis after power is turned on Turn it on in advance of starting analysis Switch on the POWER on the left side of the Main Unit The Start Up screen will be displayed and touch anywhere of screen to go to Main Menu Power switch On Turning on automatically from Sleep Mode...

Страница 48: ... function is enabled below screen will appear instead of the Start Up screen after turning on To log in touch each USER and PASS column Then input User ID and Password corresponding to the configuration from displayed key board Note that the input is case sensitive Current versions of programs Main Control and Detect USER column versions of programs PASS column versions of programs ...

Страница 49: ...d with chemicals or tap water Accurate analysis results may not be obtained if the detergent is not prepared correctly When pouring detergent into the Wash Tank take care to avoid touching the tubes that enter the Wash Tank with your bare hands This Analyzer will stop sampling when the level of the Wash solution becomes lower than a certain level In that case refill Wash solution and touch Start o...

Страница 50: ...ng the Waste Tank This Analyzer will not stop when the Waste Tank becomes full Be sure to empty the Tank before starting analysis Waste should be handled as medical waste and disposed of by recognized laboratory practices Waste in the Waste tank should be discarded Caution ...

Страница 51: ...nalysis results may not be obtained if the Reagent is not prepared correctly Do not mix the Reagent from bottle to bottle in any case Setting the Reagent in place Handling Precautions Prepare sufficient volume of Reagent for the analysis to be performed 1 Let the Reagent warm to room temperature and then mix it well and gently by repeated inversion Handling Precautions Mix the Reagent by repeatedl...

Страница 52: ...Bottle Lid on to prevent evaporation condensation or contamination Open the Lid before starting measurement Setting Reagent volume The Reagent volume must be revised when the Reagent Bottle is replaced To revise the volume touch the Latex column and input the volume of the Reagent from the numeric keypad 1 Touch the Latex column A numeric keypad will appear Reagent Bottle Lid Latex column Numeric ...

Страница 53: ... 4 2 Using the numeric keypad input the Reagent volume mL 3 Press ENTER key to register The number of available test count will be calculated from the registered volume and displayed in the Latex column The initial volume of each Reagent Vials is 18 mL Note ...

Страница 54: ...d if the Buffer is not prepared correctly Handling Precautions Prepare a sufficient volume of Buffer for the measurement to be performed This Analyzer requires approximately 190ul of Buffer per measurement 1 Let the Buffer warm to room temperature and then mix it gently by repeated inversion Handling Precautions Mixing of the Buffer should be done gently If air bubbles mix into the Buffer accurate...

Страница 55: ...e into the hole of the Bottle Head then lifting the Buffer Bottle together with the Bottle Head into the Main Unit set the bottle gently into its place 5 Align the bottom of the Buffer Cover with the guide to attach it and then press on the mark to lock the cover to close it Right side of the Main Unit Bottle Head Buffer Tube Buffer Bottle Right side of the Main Unit Bottle Head Buffer bottle Righ...

Страница 56: ...dling Precautions After setting or replacing a Buffer Bottle touch the Wash icon on the Main Menu screen to run a wash The Wash action removes air bubbles from inside the tube Do not touch the Buffer Tube by hand or Latex Glove to prevent contamination of Reagent ...

Страница 57: ...d each Cell Cassette by its handle and align its groove on the bottom to the rail on the center of the Cell Cassette Table Then slide the Cassette into place The Cell Cassette Table has R and L sides so place a Cell Cassette onto the rail of each side 2 Slide the Cell Cassette gently to the back of the Cell Cassette Table so that their guide groove fit in the projections at the rear Pushing a Cell...

Страница 58: ...Operation Manual Analysis Operations 5 13 Rev 4 In case the first picking up position is different refer to 4 1 5 Cell Set for setting the first position of picking up the Cell Correct positioning of Cell Cassette ...

Страница 59: ...ve screen displayed scan the barcodes When the Reagent lot is changed register the barcodes on the Master Curve card provided with the EXTEL HEMO AUTO HS Latex in order from the top The Reagent lot number the expiration date and the Master Curve parameters from 1 to 7 are input sequentially When the Calibrator lot is changed register the barcodes on the Calibrator Card provided with the EXTEL HEMO...

Страница 60: ...ate calibration result may not be obtained if the Calibrator is not prepared correctly Use new and clean Sample Cup Using Sample Cup which is not clean may cause incorrect result of calibration and control Definition of Racks Racks are assigned exclusively to Calibrator Control and Sample Collection Picker They can be identified with the colors and labels White Rack with Red labels STD for Calibra...

Страница 61: ... and S2 is for Calibrator H Handling Precautions The cups with Calibrators cannot be identified visually If they are placed to wrong positions correct calibration result cannot be obtained Preparation of Control 1 Prepare the two concentrations of Control and dispense the necessary volume from the each vial into a sample cup using a pipette Refer to The Instructions for Use for how to prepare Cont...

Страница 62: ...opposite to the cap Handling Precautions If bubbles remain at the bottom the cap side or midway in the Sample Collection Picker accurate measurement values might not be obtained If the seal covering the top of the Picker is exposed to possible contaminant accurate measurement values might not be obtained In this case peel the seal off before measuring 2 Set each Sample Collection Picker to the Sam...

Страница 63: ...n has to be selected to set the orders There is no special instruction required to Auto measurement mode but Analyzer recognizes the Adaptor for presence of Cups Remove the unnecessary Adaptors prior to measurement Handling Precautions If bubbles remain at the bottom the cap side or midway in the Sample Collection Picker accurate measurement values might not be obtained The Adaptors should be plac...

Страница 64: ...ng Racks to the Analyzer 1 Loading Racks Place a Rack to the Handle side and slide so that the Notch fits into the Guide Plate 2 Setting the Sample Tray onto the Main Unit Place the front tip of the Tray on the Tray position slide forward so that bottom of the Handle side comes to the top of White Stopper Then push it down so that the bottom surface completely fits to the Tray position and the Tra...

Страница 65: ... White Stopper Bottom surface is completely fitting to the Tray position and held by White Stopper The same level between the Tray and transport is achieved Tray Presence Sensor Bottom of Handle side is placed on the top of White Stopper Push it down Lower angle Lower Angle ...

Страница 66: ...lysis Touch the Start icon on the Order screen to start analysis Analysis screen will be displayed on the screen 2 Reporting results During analysis the results will be printed out real time Adding samples during analysis When all the sampling finished chimes ring and the message Remove I Tray for Sample addition appears Racks can be added or replacing Tray followed by this message and then press ...

Страница 67: ... given The result with the flag is usually positive but it is necessary to check the Sample Collection Picker if feces is too much inside or not If so it might cause the phenomena and the result is considered invalid Recollection of sample should be preferred B Reagent Blank error The Analyzer checks the IST before adding sample The flag appears when the value does not fall in the range set as nor...

Страница 68: ...ace the Buffer bottle with a bottle filled with distilled water or equivalent solution before performing this process Touch Main Menu icon then touch Wash on the Main Menu 2 Putting away the Reagent and Samples Once analysis is complete discard all the used Cells in the Cell Waste Box and Samples on the Sample Rack Tray Remove the Reagent from the Main Unit put the cap back and store them in a che...

Страница 69: ...he next start up is planned more than one week later it is recommended to switch off the Analyzer and cover it to prevent dust come in Sleep Mode Touch Sleep Mode button at the bottom of the Main Menu Touch the column for date to move the date forward and the time columns to input the time from the numeric keypad for the Analyzer to restart automatically Then touch Sleep Mode NOW button to enter S...

Страница 70: ... with a right angle 3 A beep can be heard and the lamp on the scanner flashes once each time when the barcode is read successfully 4 When all the required reading is complete a confirmation window will appear Touch OK 5 Check that the registered information is displayed on the screen 6 Touch the RETURN button to exit from barcode input mode Before registration After registration Placing the Reader...

Страница 71: ...for Master Curve Changing parameters other than mentioned below especially Cut off T0 Min IST and T7 Max IST may lead to unexpected diagnostic errors Never change parameters other than the ones mentioned below Clearing input protect In order to input data and information manually it is required to clear input protection following the process starting from the System menu 1 Touch System button on t...

Страница 72: ...data for Calibrators to be input No 1 STD Revise Conc1 Conc1 for Calibration Curve No 1 No 1 STD Revise Conc2 Conc2 for Calibration Curve No 1 No 2 STD Revise Conc1 Conc1 for Calibration Curve No 2 No 2 STD Revise Conc2 Conc2 for Calibration Curve No 2 Touch the corresponding column Then input the value with the displayed numeric keypad in accordance with the Calibrator Card 4 Go to Calibration sc...

Страница 73: ...g input protection touch Calibration on the Main Menu 2 Touch Master button at the bottom of the Calibration screen 3 Touch Exp Edit button at the bottom of the Calibration Master Edit screen 4 A numeric keypad appears Input Expiration information MM YY referring to the Master Curve Card and touch ENTER 5 After input confirm that the input data are correct according to the Master Curve Card 6 Afte...

Страница 74: ...ut Lot number referring to the Master Curve Card and touch ENTER 5 After input confirm that the input data are correct according to the Master Curve Card 6 After confirmation touch Clear column in the User Parameter screen to change it back to Protect to avoid any unexpected change ...

Страница 75: ...e first position of picking up the Cell The Analyzer memorizes which Cell on the Cassettes to pick up first even when the main power is switched off Therefore it is not required to take care of Cells as long as the Cassettes replaced timely If required refer to 4 1 5 Cell Set ...

Страница 76: ...pen it 2 Insert the printer paper as seen in the picture drawing below 3 Set the printer paper with the end slightly protruding from the paper cover Handling Precautions Note that the upper and under side of the printer paper should be distinguished Set the paper so that the rounded side of the roll faces to the surface 4 Pull out around 3cm of the printer paper 5 Close the paper cover to its orig...

Страница 77: ...Operation Manual Analysis Operations 5 32 Rev 4 6 Insert the printer paper that is pulled out into the printer paper guide take the paper out of the guide first before cutting it Printer paper Printer paper guide ...

Страница 78: ...ing Emergency STOP icon To return to the normal operation take actions below 1 Check that the Analyzer has completely stopped 2 Turn the power switch OFF to switch the Analyzer off 3 Wait for at least 10 seconds then switch the Analyzer on again 4 Touch the Home Position icon to start initialization If the above steps do not fix the error contact local distributor ...

Страница 79: ... Implementing troubleshooting activity may include exposure to danger of biohazard and sharp edge Wear rubber globe and take extra care to avoid infection and injury 6 1 Structure of the Codes The error codes consist of five digits of numbers The first two digits represent the status of the Analyzer at the time of the error The last three digits represent the description of the error If necessary ...

Страница 80: ... Unit Control operation 11 Adjustment Test 12 Analysis for Calibration Curve Creation 13 Transfer of PPMC configuration data 14 Transfer of temperature control parameter 15 Switching Temperature control mode between Stand by and Ready 16 Reception of temperature control parameter 17 Detect Test 18 Temperature Control CPU version check 19 User Maintenance 20 Moving Cell Cassette to the original pos...

Страница 81: ...ction table index sensor error 009 Right cell cassette error 065 Error Reaction table is empty 010 Left cell cassette error 066 Error Waste Tank is full 011 Sample syringe error 067 Error Wash Tank is empty 012 Reagent syringe error 069 Error Device cover is open 013 Reaction table error 070 Error Latex cover is closed 021 Reagent switching valve error SV1 077 Cell Hand index sensor error 022 Wash...

Страница 82: ...ng maximum racks Please wait until end of current run 104 Transportation rack origin sensor error 205 Reaching maximum racks Please wait until end of current run 105 Import rack origin sensor error 206 LIS communication error 106 Cell Hand horizontal origin sensor error 208 Error Movement time out 107 Cell Hand Up Down origin sensor error 209 Error CONTROL CPU Not receive the command 108 Right cel...

Страница 83: ...ribution conflict 308 Barcode n error 238 Order attribution conflict 309 Barcode nnn error 239 Control order conflict 310 Barcode hhh not Hex 240 Latex is empty 311 Barcode hh FF 241 Latex level is low 312 Barcode hhh aaa Over 242 Buffer is empty 313 Barcode aaa error 243 Buffer level is low 314 Barcode EEP_ROM broken 244 Latex Buffer is empty 315 Barcode initialize memory error 245 Latex Buffer l...

Страница 84: ...d the limit USB 350 Master Curve registration has been already finished 387 The attribution of the file cannot be found in the USB memory 351 Master Curve Lot Exp Data error 388 No date can be found in the result file in the USB memory 352 Master Curve Conc IST data error 396 System is in process 353 Calibrator registration has been already finished 397 System is in process 354 Calibrator Lot Exp ...

Страница 85: ...sticky If the conditions are clear perform Home Position Errors related to Cell Transport This is the error related to Cell transportation If this error occurred check the conditions below The Cell Cassettes are set properly The Cell Waste Box is placed correctly and capable No cell is dropped in the range of cell transportation Switch off the Analyzer and Remove Cell Hand Cover Then visually chec...

Страница 86: ...correctly Barcodes for Sample should be as vertical as possible and the Scanner requires approximately 4mm of blank space on the upper and lower side of Barcodes The quality of the Barcode is fine The shape of the Barcode is clear not scratched and not blurred Format of Barcode is acceptable Acceptable format is defined in the Communication Parameter If they do not match the Barcode information ca...

Страница 87: ...e B as specified by JIS T1001 Hemoglobin Latex agglutination reaction method Integrating Sphere Turbidity 200 samples hour 1 item 7 400ng mL 10 60 L sample metering pump Latex 20 300 L Buffer 20 300 L 300 L maximum Number of samples able to be set at one time 80 10 samples rack 8 racks Stool sample containers intended for the purpose Method of direct insertion of stool sample container with mixing...

Страница 88: ...HM JACKarc Operation Manual Miscellanea 7 2 Rev 4 7 2 Outline Dimensional Drawing Front Left side Right side Back Top ...

Страница 89: ... 3 Rev 4 7 3 Communication Specification This Analyzer is equipped with communication functions to enable output of analysis results to a host computer If Communication Specification is required refer to HM JACKarc Communication Specification ...

Страница 90: ...1405 1 coupling 4 E855707 1 joint 5 HSL369025 synapse SM5 1020L 6 HSL369112 mini fitting VFI346 2 HSL312703 mini fitting VFT306 1 HSF275003 Teflon tube φ1 φ2 HSG153006 Tygon tube R3606 1 8 1 4 Unitube 6 φ6 φ8 5 SAMPLE sample syringe Injection arm REAGENT reagent syringe WASH wash syringe Buffer fluid bottle Waste water pump R wash tank S wash tank Water supply tank Waste water tank ...

Страница 91: ...rigin position photo interrupter Position sensor photo interrupter Sample rack presence sensor Sample container presence sensor Cell transportation Hand vertical stepping motor Hand horizontal stepping motor Cassette feed stepping motor Origin position photo interrupter Position sensor photo interrupter Reaction table Table rotation stepping motor Origin position photo interrupter Position sensor ...