9
PSC (L-N/L-L) / PFC (L-PE)
Function
Rated
Voltage
Nominal Test Current
at 0
Ω
External Loop:
Magnitude/Duration(*5)
Range
Accuracy
PSC
100~500V
50/60Hz
6A/20ms
PFC
100~260V
50/60Hz
6A/20ms
2A/20ms
15mA/500ms
PFC
(ATT)
100~260V
50/60Hz
L-N: 6A/60ms
N-PE: 10mA/approx. 5s
2000A/20kA
Auto-Ranging
PSC/PFC
accuracy is
derived from
measured loop
impedance
specification and
measured voltage
specification
*5: at 230V
RCD
Accuracy
Trip Current
Function
Rated Voltage
AC Type
A Type
Trip Time
X1/2
-8%~-2%
-10%~0%
X1
+2%~+8%
0%~+10%
X5
+2%~+8%
0%~+10%
Ramp(
)
230V+10%-15%
50/60Hz
±4%
± 10%
±(1%rdg+3dgt)
Auto
Depending on the accuracy at each function.
Measurement sequence:
X1/2 0°→X1/2 180°→X1 0°→X1 180°→X5 0°→X5 180°
Measurements with x5 are not carried out for RCDs with
nominal current of 100mA or more.
RCD(Uc)
Function
Rated Voltage
Range
Test Current
Accuracy
UC
230V+10%-15%
50/60Hz
100.0V
1/2I n
(max150mA)
+5% +15%rdg
±8dgt