10
RCD Trip Current Duration
RCD Trip Current Duration
Function
Type
10
30
100
300
500
1000
AC
2000
2000
2000
2000
2000
2000
G
A
2000
2000
2000
2000
2000
n.a
AC
2000
2000
2000
2000
2000
n.a
X1/2
S
A
2000
2000
2000
2000
2000
n.a
AC
550
550
550
550
550
550
G
A
550
550
550
550
550
n.a
AC
1000
1000
1000
1000
1000
n.a
X1
S
A
1000
1000
1000
1000
1000
n.a
AC
410
410
410
n.a
n.a
n.a
G
A
410
410
410
n.a
n.a
n.a
AC
410
410
410
n.a
n.a
n.a
X5
S
A
410
410
410
n.a
n.a
n.a
AC
n.a
G
A
Goes up by 10% from 20% to 110%
300ms
×
10 times
n.a
AC
n.a
Trip
Current
Duration
(ms)
Ramp
(
)
S
A
Goes up by 10% from 20% to 110%
500ms
×
10 times
n.a
Earth
Measuring
Frequency
Range
Accuracy
20Ω range : ±(3%rdg+0.1dgt)
200/2000Ω range : ±(3%rdg+3dgt)
825Hz
20/200/2000Ω
Auto-Ranging
(Auxiliary earth resistance 100±5%)
PHASE ROTATION
Rated Voltage
Remarks
50-500V
50/60Hz
Correct phase sequence: are displayed “1.2.3” and mark
Reversed phase sequence: are displayed “3.2.1” and mark
Volts
Function
Rated voltage
Measuring
Range
Accuracy
Volts
25~500V
45~65Hz
25~500V
±(2%rdg+4dgt)
Frequency
25~500V
45~65Hz
45~65Hz
±(0.5%rdg+2dgt)