5.1
front view
1
2
3
4
1. Laser aperture.
To clean the laser aperture window from inside or to swap the
aperture plate for different type of aperture (i.e. DiscoScan
lens attachment bracket), remove the six socket bolts that
hold the laser aperture in place and remove it.
2. Aperture masking plate.
This metal masking plate can be moved left and right when
locking bolt is loosened.
3. Attachment bracket.
The attachment bracket spins 360 degrees around the
laser body whilst it can be locked in desired position by two
small locking handles on sides. Due to the sophisticated
locking mechanism there’s no need to use excessive force
when locking the system in position.
4. Laser emission indicator.
When this indicator is lit up the laser system is ready to emit
the laser radiation as soon as it receives instructions from
control software.