function and set these two constant.
How to use MX+B function
You can use MX+B function from the front panel operation or the remote
interface operation.
Front Panel Operation
To set the values of M and B, user needs to configure this math function
by pressing CONFIG and SHIFT + %. Use
◁
and
▷
to switch between M
and B. Press ENTER on your selection. Then use
◁
and
▷
to move
through the digits and use
△
and
▽
to increase or decrease the number
to a desired value. Press ENTER to set the value. User needs to set the
values of M and B first. Then Select a measurement function and press
SHIFT + % to activate MX+B function as shown as
Figure 4-20
.
Figure 4-20
Remote Interface Operation
Use the following commands to enable and configure MX+B function:
CALCulate:FUNCtion MXB
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:MXB:MMFactor {<
value
>|MINimum|MAXimum}
CALCulate:MXB:MMFactor? [MINimum|MAXimum]
CALCulate:MXB:MBFactor {<
value
>|MINimum|MAXimum}
CALCulate:MXB:MBFactor? [MINimum|MAXimum]
Содержание M3500A
Страница 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Страница 2: ...Operation Manual M3500A DMM ...
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 26: ...1 2 3 4 5 6 ...
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 47: ......
Страница 48: ...48 Figure 3 18 5 ...
Страница 95: ......
Страница 104: ......
Страница 151: ......
Страница 155: ......