background image

 

Field Procedures

 

1

 

17

th

 Edition, 2015 

 

MULTIMETER: YSI-ProDSS  

rev. 12/16/15 

EQUIPMENT NEEDED 

 

YSI-ProDSS multimeter 

 

data sheet, clipboard, pencil 

Note:

  

The battery may be damaged if kept in 

an environment outside the range of -20 - 60°C 
(-4° - 140°F). 

NOTE 2015-16:  

We purchased one instrument in 

summer 2015 in the expectation of purchasing a 
second and making it our standard field water-
quality testing equipment in summer 2016.  
Meanwhile, we are experimenting with settings 
and field instructions, and recording values side-
by-side with our current equipment. 

INITIAL INSTRUMENT PREPARATION: 

1.

 

CABLE CONNECTION 

Inspect the cable connection between the 
handheld meter and the cable.  If cable is not 
connected to the meter, 

make sure the 

connectors are clean and dry before 
connecting!

   

If connecters are clean and dry, 

align the keys on the cable assembly connector 
with the slots on the instrument connector. Push 
together firmly, and then 
twist the outer ring 
clockwise until it locks 
into place.  The metal pin 
should “snap” into place 
and be visible through 
small hole in the outer 
ring (see image).   

2.

 

SENSORS AND GUARD CHECK 

If any of the individual sensors are not 
connected, refer to “Settings and 
Troubleshooting” section for sensor installation.  
If there is an empty “slot” with no sensor 
installed, 

a port plug or sensor must be 

installed or the bulkhead will be damaged! 

1.

 

Remove calibration cup (see diagram):  firmly 
hold the calibration cup with one hand and 
twist the cup’s retaining nut counterclockwise 
until it loosens (don’t keep turning; once it is 
loose you can slide it off).  Remove the cup 
and set aside in a clean, dry place. 

2.

 

Firmly hold the bulkhead with one hand and 
turn the sensor guard counter-clockwise until 
completely unthreaded.  Set aside the sensor 
guard in a clean dry place.  

Use caution 

once the sensors are exposed!

 

3.

 

Conduct a quick visual inspection of the 
sensors; clean any debris on the sensor tips 
with a lint-free cloth and rinse with water

.

  

Carefully “test” the connection of each sensor 
by grasping the sensor near the bulkhead 
between your thumb and finger, and try to 
turn the sensor to determine if it is loose.  If 
the sensor moves or twists 

at all 

(sensors 

should 

firmly

 be in place), reconnect the 

sensor guard and inform the team leader.  
(See “Settings and Troubleshooting,” below, 
under “Sensor Installation”.)  If sensors are 
firmly in place, proceed to next step. 

4.

 

Carefully slide the sensor guard over the 
sensors and push it toward the bulkhead until 
the sensor guard threads align with the 
bulkhead threads.  Carefully finger-tighten 
the sensor guard clockwise.  

NOTE:

  

If any resistance is felt, loosen the 

sensor guard completely to prevent cross-
threading. Incorrect installation may cause 
damage to the sensor guard or bulkhead. 

Do not 
remove 
this ring! 

Simply 
loosen 
the 
ring to 
slide 
the 
entire 
cup off. 

Summary of Contents for ProDSS

Page 1: ...st be installed or the bulkhead will be damaged 1 Remove calibration cup see diagram firmly hold the calibration cup with one hand and twist the cup s retaining nut counterclockwise until it loosens d...

Page 2: ...e battery back up to 80 Expected battery life 20 hours at 25 default LCD brightness 14 hours at 100 LCD brightness Study the main Run screen Confirm the following The battery charge is 10 If not infor...

Page 3: ...oxes are checked Sample Period s is set to 15 Sample Count is set to 10 c Stabilize Hit the up arrow and select Start Auto Stable You ll return to the Run screen On the Run screen there will be a flas...

Page 4: ...f your pairs of readings are within the acceptable precision limits resample as needed 9 Record time and full initials of the sampler this should be someone who has been properly trained see Quality A...

Page 5: ...ination If the port is wet carefully dry with a lint free cloth 2 Apply a thin coat of lubricant to the sensor o rings Wipe off any excess lubricant with a lint free cloth 3 Carefully align the sensor...

Page 6: ...and let dry before replacing the port cover Battery pack troubleshooting If the instrument is not turning on do the following 1 Attempt to charge the unit Using YSI charging equipment only plug unit i...

Page 7: ...he logging settings are inaccurate press the system button scroll down to the logging submenu option go to the Continuous Mode and uncheck the box by pressing Enter The box should become unchecked Pre...

Page 8: ...This page left blank for printing...

Reviews: