Front Panel Operation
Before using the reading memory feature, user needs to select a
measurement function (or the math function) first and then select the
trigger mode.
To store readings, use the following steps:
Press STORE button, and the multimeter will start to store the readings
produced until the specified number of readings is reached. The default
number of readings to store is 100. To change this number, press
CONFIG and then STORE buttons. Then use
◁
and
▷
buttons to move
between the digits and
▽
and
△
buttons to increase or decrease
numbers as desired. When ready, press ENTER button.
The default
number of readings will be restored when the meter has been turned off.
The location of the buttons are shown with red rectangle frames in
Figure 4-28
.
※
Note: The MEM annunciator will be lit while the multimeter stores
readings and be turned off when the specified number of readings is
reached.
Procedures: STORE
→
<NUMBER>
Figure 4-28
To recall the stored readings, use the following steps:
Press RECALL button, and the multimeter will display all the stored
readings starting from the first reading. Use
◁
and
▷
or
▽
and
△
Содержание M3500A
Страница 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Страница 2: ...Operation Manual M3500A DMM ...
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 26: ...1 2 3 4 5 6 ...
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 47: ......
Страница 48: ...48 Figure 3 18 5 ...
Страница 95: ......
Страница 104: ......
Страница 151: ......
Страница 155: ......