Figure 4-4
Remote Interface Operation
The automatic input resistance mode can be enabled / disabled. With
AUTO OFF (default), the input resistance is fixed at 10M
Ω
for all ranges.
With AUTO ON, the input resistance is set to >10G
Ω
for the three lowest
DC voltage ranges (100mV, 1V and 10V). Use the following command
from your PC terminal to disable the auto input DC resistance setting
(the result is a fixed input DC resistance at 10M
Ω
for all
measurements.)
INPut:IMPedance:AUTO {OFF | ON}
4.1.5 Threshold Resistance (Continuity)
Definition
When testing continuity, the beeper goes off when the measured
resistance is less than the threshold resistance. The threshold resistance
can be set to any value between 1
Ω
and 1000
Ω
.
Default
The factory default for continuity threshold resistance is 10
Ω
. User’s
selection is stored in a volatile memory and the default value will be
restored after the meter has been turned off.
How to set the threshold resistance
User can change the threshold resistance only through the front panel.
Содержание M3500A
Страница 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Страница 2: ...Operation Manual M3500A DMM ...
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 26: ...1 2 3 4 5 6 ...
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 47: ......
Страница 48: ...48 Figure 3 18 5 ...
Страница 95: ......
Страница 104: ......
Страница 151: ......
Страница 155: ......