13
OMH
100mV
100mV
400
W
4000
W
Max, output
2mA
-100.00~1100.00mV
0.0~400.0
W
0.0~4000.0
W
0.1mV
0.1
W
1
W
0.05%+0.3mV
0.05%+0.2
W
0.05%+2
W
+0.5~+3mA
Note 1,Note2
+0.05~+0.3mA
Note 1,Note2
-
-
-
-
12
Section Six Instrument Measurement
Warning
During the operation, never apply more than 30V between
any two terminals, or between any terminal and earth ground.
Any voltage more than 30V will not only do damage to the
Instrument but also lead to possible personal injury.
!
Caution
•
During the operation, never apply more than 30V between any
two terminal, or between any terminal and earth ground. Any
voltage more than 30V will not only do damage to the Instrument.
but also lead to possible personal injury.
•
When connecting to the Instrument, the power supply of the
device under test should be cut off. Otherwise, any connection
with a device without cutting off its power supply will cause
possible damage to the Instrument.
Input Operation Procedure
Function
Operation
n
%
Operation
Display
Measurement
Range
DCI
20mA
00.000mA
-025.00 mA%
0.000 V
-1.000~22.000
mA
-31.25~112.50
mA%
-0.2000~28.000 v
DCV
28 V
20mA
%
!
Measuring DC Current
1) Insert one end of the test lead into the mA jack of the
Instrument (INPUT) terminal and connect the other end to the
output of the user’s device as shown in Figure 5-3;
2) Press the key (OUTPUT / IN), the LCD displays ‘INPUT’ and it
denotes that the Instrument is in an input state.
3) Press the key (mA%/V) to select the input to be set in ‘mA’ or
‘mA%’ and the LCD displays unit ‘mA’ or ‘mA %’in which the
value of 0% denotes 4mA and the value of 100% denotes
20mA.
4) The Instrument starts measurement, and the LCD displays ‘ON’
and the measured result simultaneously.
5) The refreshing rate of ,measurement result is twice every
second. And the LCD displays ‘OL’ if the measured value
exceeds the measuring range.
Figure 5-3