4) Set the Standard source to an appropriate range;
5) Set the output of standard source as the value given in Table
10-3 and when the output stabilized, press the (100%/START) key
and the LCD displays ‘SAVE’ denoting that the calibrated point
has been stored.
6) Press the key (STEP/AUTO) and the LCD displays ‘CAL FS’ ,
and repeat the operation of steps 5
Voltage Measurement Calibration
1) Press the key (mA % / V) to select voltage measurement
function , and the LCD displays ‘ INPUT ’ CAL 0 V;
2) Connect to the Instrument as shown in Figure 10-3;
3) Set the standard source to an appropriate range.
4) Set the output of standard source as the value given in Table
10-3 and when the output stabilized, press the key
(100%/START), and the LCD displays ‘SAVE’ denoting that the
calibrated point has been stored.