TRMS

DIGITAL CLAMPMETER

MODEL - KM 078

I. SAFETY :

This  manual  contains  information  and  warnings  that  must  be 
followed  for  operating  the  instrument  safely  and  maintaining  the 
instrument in a safe operating condition. If the instrument is used in
manner not specified by the manufacturer, the protection provided
by the instrument may be impaired.

The meter protection rating, against the users, is double insulation
per  UL/IEC/EN61010-1  Ed.  3,  IEC/EN61010-2-033  Ed.  1,  CAN/
CSA C22.2 No. 61010-1 Ed. 3, IEC/EN61010-2-032 Ed. 2 & IEC/
EN61010-031 Ed. 1.1 :
Measurement Category III 600V & Category IV 300V AC & DC.

Per IEC610-1 (2010) OVERVOLTAGE CATEGORY

OVERVOLTAGE CATEGORY II (CAT II)

 is for equipment intended 

to  be  supplied  from  the  building  wiring.  It  applies  both  to  plug-
connected  equipment  and  to  PERMANENTLY  CONNECTED 
EQUIPMENT.

OVERVOLTAGE CATEGORY III (CAT III)

 is for equipment intended 

to  form  part  of  a  building  wiring  installation.  Such  equipment 
includes  socket  outlets,  fuse  panels,  and  some  MAINS  installation
control equipment.

OVERVOLTAGE CATEGORY IV (CAT IV)

 is for equipment installed 

at  or  near  the  origin  of  the  electrical  supply  to  a  building,  between
the  building  entrance  and  the  main  distribution  board.  Such 
equipment  may  include  electricity  tariff  meters  and  primary  over
current protection devices.

Terms in this manual :

WARNING

Identifies  conditions  and  actions  that  could result in 
serious injury or even death to the user.

CAUTION

Identifies   conditions  and  actions   that  could  cause
damage or malfunction in the instrument.

01

Содержание KM 078

Страница 1: ...with DC Zero mode on DCA DC ACA ranges Temperature Coefficient Nominal 0 15 x o o o o o specified accuracy C 0 C 18 C or 28 C 40 C or otherwise specified Power Supply Standard 1 5V AAA Size Battery X...

Страница 2: ...M AMPTIP CLAMP ON DC CURRENT 1 2 3 Accuracy Resolution 2 0 rdg 5dgts 0 01 A 60 00 A 1 Induced error from adjacent current carrying conductor 0 01A A 2 Specified with DC Zero mode applied to offset the...

Страница 3: ...00 0 433 0 416 3 8 2 000 0 331 0 243 26 5 2 000 0 242 0 130 46 2 One limitation to making true RMS measurements is crest factor and you should consider crest factor when making AC measurements Crest f...

Страница 4: ...Thermo Anemometer Thermo Hygrometer Distance Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Ca...

Страница 5: ...s for equipment intended to be supplied from the building wiring It applies both to plug connected equipment and to PERMANENTLY CONNECTED EQUIPMENT OVERVOLTAGE CATEGORY III CAT III is for equipment in...

Страница 6: ...I 600V or better This Clamp on meter is designed to apply around or remove from uninsulated hazardous live conductors But still individual protective equipment must be used if hazardous live parts in...

Страница 7: ...cessories Test leads set User Manual Soft carrying case V SPECIFICATIONS 05 Safety Double insulation per UL IEC EN61010 1 Ed 3 IEC EN61010 2 033 Ed 1 CAN CSA C22 2 No 61010 1 Ed 3 IEC EN61010 2 032 Ed...

Страница 8: ...dg 5dgts 10 mA 50Hz 60Hz Diode Tester Test Current 0 3mA typically Open Circuit Voltage 3 5VDC typically 2 000 V 1 5 rdg 5dgts 1 Accuracy Resolution Range 1 mV 07 1 Induced error from adjacent current...

Страница 9: ...ep tones proportional to the field strength Detection Frequency 50 60Hz Detection Antenna Inside the top side of the stationary jaw Probe Contact EF Detection For more precise indication of live wires...

Страница 10: ...DCA polarity at where best current accuracy is specified 11 VII OPERATION WARNING Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to de...

Страница 11: ...I 600V CAT IV 300V PEAK RMS REC SELECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V EF Live Ground 13 CAUTION WARNING for Non Invasive Clamp on Curr...

Страница 12: ...HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V 6008 Hz SELECT 1 HZ Non Invasive Regular Current Functions ACA DCA DC ACA Hz Input is made via the clam...

Страница 13: ...ECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V 0602 W V AUTO PEAK RMS REC SELECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A...