Note :

 When using diode test function, normal forward voltage drop

(forward  biased)  for  a  good  silicon  diode  is  between  0.400V  to 
0.900V. A reading higher than that indicates a leaky diode (defective).
A zero reading indicates a shorted diode (defective). An OL indicates 
an  open  diode  (defective).  Reverse  the  test  leads  connections 
(reverse biased) across the diode. The digital display shows OL if the
diode  is  good. Any  other  readings  indicate  the  diode  is  resistive  or
shorted (defective).

CAUTION :

Using  Resistance,  Continuity  or  Diode  function  in  a  live  circuit  will
produce  false  results  and  may  damage  the  meter.  In  many  cases 
the  suspected  component(s) must be disconnected from the circuit
to obtain an accurate measurement reading.

W

 Resistance,      Continuity &      Diode functions :

Inputs    are    made    through    the    test    lead    terminals   

COM/+

Defaults  at  

W

 Resistance

.  Press  

SELECT 

 button momentarily to

select the subject functions in sequence.

0l

.

V

AUTO

PEAK-RMS

REC

SELECT

HOLD

ZERO

OFF

MAX 600A

600V CAT III
300V CAT IV

EF

W

A

Hz

Hz

V

+

A

Hz

COM

+

CAT III

600V

CAT IV

300V

+

+

0602

.

W

V

AUTO

PEAK-RMS

REC

SELECT

HOLD

ZERO

 

OFF

MAX 600A

600V CAT III
300V CAT IV

EF

W

A

Hz

Hz

V

+

A

Hz

COM

+

CAT III

600V

CAT IV

300V

+

+

SELECT

1

 

0005

.

W

AUTO

PEAK-RMS

REC

SELECT

HOLD

ZERO

 

OFF

MAX 600A

600V CAT III
300V CAT IV

EF

W

A

Hz

Hz

V

+

A

Hz

COM

+

CAT III

600V

CAT IV

300V

+

+

SELECT

1

3802

.

k

W

AUTO

PEAK-RMS

REC

SELECT

HOLD

ZERO

 

OFF

MAX 600A

600V CAT III
300V CAT IV

EF

W

A

Hz

Hz

V

+

A

Hz

COM

+

CAT III

600V

CAT IV

300V

+

+

W

Inputs are made through the test lead terminals 

COM/+

. Defaults at

 

   Capacitance

Capacitance functions :

CAUTION :

1. Using  Capacitance  function  in  a  live  circuit  will  produce  false 

results  and  may  damage  the  meter.  In  many  cases  the 
suspected  component(s) must be disconnected from the circuit
to obtain an accurate measurement reading.

2. When using Capacitance function, discharge capacitor(s) before

making  any  measurements.  Large  value  capacitors  should  be
discharged through an appropriate resistance load.

i536

m

F

AUTO

PEAK-RMS

REC

SELECT

HOLD

ZERO

OFF

MAX 600A

600V CAT III
300V CAT IV

EF

W

A

Hz

Hz V

+

A

Hz

COM

+

CAT III

600V

CAT IV

300V

+

+

+

17

Содержание KM 078

Страница 1: ...with DC Zero mode on DCA DC ACA ranges Temperature Coefficient Nominal 0 15 x o o o o o specified accuracy C 0 C 18 C or 28 C 40 C or otherwise specified Power Supply Standard 1 5V AAA Size Battery X...

Страница 2: ...M AMPTIP CLAMP ON DC CURRENT 1 2 3 Accuracy Resolution 2 0 rdg 5dgts 0 01 A 60 00 A 1 Induced error from adjacent current carrying conductor 0 01A A 2 Specified with DC Zero mode applied to offset the...

Страница 3: ...00 0 433 0 416 3 8 2 000 0 331 0 243 26 5 2 000 0 242 0 130 46 2 One limitation to making true RMS measurements is crest factor and you should consider crest factor when making AC measurements Crest f...

Страница 4: ...Thermo Anemometer Thermo Hygrometer Distance Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Ca...

Страница 5: ...s for equipment intended to be supplied from the building wiring It applies both to plug connected equipment and to PERMANENTLY CONNECTED EQUIPMENT OVERVOLTAGE CATEGORY III CAT III is for equipment in...

Страница 6: ...I 600V or better This Clamp on meter is designed to apply around or remove from uninsulated hazardous live conductors But still individual protective equipment must be used if hazardous live parts in...

Страница 7: ...cessories Test leads set User Manual Soft carrying case V SPECIFICATIONS 05 Safety Double insulation per UL IEC EN61010 1 Ed 3 IEC EN61010 2 033 Ed 1 CAN CSA C22 2 No 61010 1 Ed 3 IEC EN61010 2 032 Ed...

Страница 8: ...dg 5dgts 10 mA 50Hz 60Hz Diode Tester Test Current 0 3mA typically Open Circuit Voltage 3 5VDC typically 2 000 V 1 5 rdg 5dgts 1 Accuracy Resolution Range 1 mV 07 1 Induced error from adjacent current...

Страница 9: ...ep tones proportional to the field strength Detection Frequency 50 60Hz Detection Antenna Inside the top side of the stationary jaw Probe Contact EF Detection For more precise indication of live wires...

Страница 10: ...DCA polarity at where best current accuracy is specified 11 VII OPERATION WARNING Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to de...

Страница 11: ...I 600V CAT IV 300V PEAK RMS REC SELECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V EF Live Ground 13 CAUTION WARNING for Non Invasive Clamp on Curr...

Страница 12: ...HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V 6008 Hz SELECT 1 HZ Non Invasive Regular Current Functions ACA DCA DC ACA Hz Input is made via the clam...

Страница 13: ...ECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A Hz COM CAT III 600V CAT IV 300V 0602 W V AUTO PEAK RMS REC SELECT HOLD ZERO OFF MAX 600A 600V CAT III 300V CAT IV EF W A Hz Hz V A...

Страница 14: ...meter consecutive measurement with the main display displaying reading as the reference value LCD turns on Press the button for one second or more to toggle Relative Zero mode HOLD 80ms PEAK RMS mode...

Страница 15: ...e like open fuses to protect the user and the instrument Most measuring functions through this terminal might then be open circuit Such components should only be replaced by qualified technician Refer...

Страница 16: ...horized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered ne...

Отзывы: