Operating
Starting & general rules
Turn on the measuring instrument with the hardware module
KECT3. On the opening screen the device name, the logo of
the manufacturer Kurth Electronic, the firmware version, the
UDMT version, the options installed, and the Hw-Id. are
displayed. Shortly after the self-test the main menu appears.
Before starting the measurements the user can make basic
device settings in the setup. In most cases, the
measurements and settings can be selected using the menu-
driven user interface. For selecting the vertical control buttons
are used, then it is confirmed with the
SEL
button. Various
cables and test parameters are selected by pressing the
function keys
F1 - F4
. In order to facilitate and speed up the
operation, some measuring modes can be selected directly
using the function keys. To move to the previous screen, press
the
ESC
key. The measurements begin immediately after
selecting the parameters.
Copper Parameters
You can easily create new copper parameters as base for
measurements within the menu item
Copper Parameters
. As
default, the common cable Y_4x2x06 is deposited. Intended
for the measurement, the cable satisfies other parameters, by
the input of only two values, such as the cable diameter and
the resistance per meter R, the remaining copper cable
parameters are calculated from KECT3. Only the cable
manufacturer specified VF-value needs to be entered. Press
F4
Edit and jump with the left-right control keys through the
values. The value is then selected with
SEL
and changed in
the context menu on the numeric keypad. Make sure that the
new settings are saved under a changed name as the default.
Jump to Name and confirm with
SEL
- then opens the context
menu to change the name - after changing the name go back
to
Line Parameters
again with
SEL
and press
F4
Save. The
new name now appears under
Cable Types
in the selection
menu. With
SEL
the desired value is selected, which then is
marked with a hook at the right margin.
Operating Manual
KE3600
xDSL
M
ULTI
T
EST
Edit
Date / Time
Copper Parameters
a/b: 0,0 V
Y(ST)Y_4x2x04
Y(ST)Y_10x2x06
3
4
5
Y(ST)Y_4x2x06
1
Y(ST)Y_4x2x08
2
25
KURTH ELECTRONIC www.kurthelectronic.de [email protected]
Help
Help
Date / Time
Copper Test
a/b: 0,0 V
TDR
KE900 Functions
Copper Parameters
3
4
5
Digital Multimeter
1
xDSL Line Qualification
2
Clear
Save
Date / Time
Copper Parameters
a/b: 0,0 V
Name
Ø
mm
0.600
R' 0.0636
W
/m
2
k 55.600 m/
W
mm
V 200.0 m/µs
VF 0.66
-1
a 0.0039 K
Help
Y(ST)Y_4x2x06
2
A 0.503 mm
2R' 0.0716
W
/m
C' 0.098 nF/m
V/2 100.0 m/µs
T 20 °C
0