The device is mainly for use in X-ray exposure rooms and hospital wards. To use it in other places, consult
our sales representative or a distributor.
Be sure to use and store this equipment under the conditions described below:
Parameter
Factor
Value
Units
AC/DC Adapter setting
(when supplying from the
output source of Adapter)
Voltage
+/-10% of rated
voltage
Vac
Frequency
50/60
Hz
Environmental
Pressure
Operation
700 to 1060
hPa
Storage/Ship
500 to 1060
hPa
Humidity
Operation
20 to 75
% RH
Storage/Ship
10 to 95
% RH
Temperature
Operation
10 to 40
°C
Storage/Ship
-15 to 55
°C
Contact duration by
operator
10 s ≤ t < 1 min
Operator Accessible Part
Enclosure of Power Supply and X-ray Detector
Contact duration by
Patient
1 min ≤ t < 10 min
Patient Accessible Part
Enclosure of X-ray Detector (PC Film)
Material Group
IIb
Pollution Degree
2
Overvoltage Category
II(When supplying from Power Supply)
Max. working condition
(Digital X-ray Flat Panel
Detector & Power Supply
for wired communication)
Continuous operated with fully charged battery and cycles for image capture
and transmission from Detector to PC
Max. working condition
(Digital X-ray Flat Panel
Continuous cycles for image capture and transmission from Detector to PC