Step 3.
After Step3, Click the “
” button
The X-ray condition should be set or tested beforehand, Click the “Get Bright” (The acquired Bright frame
“Bright_ image resolution_bit_bright value.raw” will be generated in the selected folder.)
Bright value [LSB] is descrived on the “table 1” and then it will be acquired 5 different level of bright images.
Step 4.
After Step3, Click the “
” button
Select the desired Dark file and 5 Bright files each, and click
.
If you want to cancel the selection, select the file and click
.
When the selection of dark and bright files is completed, click “RUN” to create the BPM file.