GENERAL SAFETY INFORMATION
KTMED Inc. Model OXYPALM
Operator’s Manual
6
※
Classification
KTMED's model OXYPALM is classified as;
- OXYPALM: CLASS IIa equipment
- OXYPALM is operated with internal battery only.
- Batteries should be replaced by a qualified person. If there is no qualified person, contact the service
center.
- BF APPLIED PART
- Ordinary equipment without protection against ingress of water
- Continuous operation
- Equipment not suitable for use in presence of a flammable anesthetic mixture by standard of EN
60601-1-1(Safety of Electric Medical Equipment)
- Equipment that complies with class B for noise emission, Level B for noise immunity, per standard of
EN 60601-1-2 (Electromagnetic Compatibility Requirements)
※
Warning
※
Caution
There is a risk of explosion if the battery is replaced with an incorrect type.
Do not:
-
dispose of the used batteries in fire;
-
put in the batteries backwards;
-
mix new and used batteries; and use other types of batteries together.
Or, there may be an explosion or a leak that can cause personal injury. Dispose of used batteries according to
the manufacture’s instructions.
The unit shall be placed far from a generator, X-ray equipment, broadcasting equipment or transmitting wires,
so as to prevent the electrical noise from being generated during the operation. When this device is near such
equipment, it can produce inaccurate measurement.