SECTION 4 – Setup
KTMED Inc. Model OXYPALM
Operator’s Manual
15
4-3-2 Alarm interval time set
Press
button about 2 seconds, and use
button to display as [Figure 14]. If you want to adjust
system sound volume, press
. You can set the alarm interval time OFF, 30, 60, 90, 120 min.
[Figure 14]
4-4 Alarm limit set
※
Default setting alarm
4-4-1 SpO2 low alarm limit set
Press
button about 2 seconds, and use
button to display as [Figure 15]. If you want to adjust
SpO2 low alarm limit, press
. You can set the SpO2 low alarm limit from 85 to 99. (1% / 1step)
[Figure 15]
4-4-2 Pulse rate high alarm limit set
Press
button about 2 seconds, and use
button to display as [Figure 16]. If you want to adjust
Pulse rate high alarm limit, press
. You can set the pulse rate high alarm limit from 35 to 255. (5bpm/
1step)
Parameter List
High limit
Low limit
SpO2 (%)
-
85
Pulse Rate(bpm)
170
40