TECHNICAL DATA
7
59
OPTISYS TUR 1060
www.krohne.com
01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en
7.2 Technical data table
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principle
90
°
scattered light method
Application range
Turbidity measurement of liquids according to ISO 7027 and US EPA 180.1
Measuring range
0...100 NTU/FNU and 0...1000 NTU/FNU factory configured
Design
The measurement system consists of a sensor and a signal converter and is only available as compact version.
Display and user interface
Display and user interface
Display and user interface
Display and user interface
Display
Multi-line liquid crystal backlight display
Operating and display language
English
Other properties
Other properties
Other properties
Other properties
Response time
Adjustable 1...60 seconds
Alarms
Two programmable, 120...240 VAC, 2A form C relay
Measuring accuracy
Reference conditions
+1...+50
°
C / +33.8...+122
°
F
95% relative humidity
Altitude up to 2000 m / 6600 ft
Accuracy
< 40 NTU/FNU:
±
2% of reading or
±
0.02 NTU/FNU whichever is greater (0...100 or
0...1000 NTU/FNU range)
> 40 NTU/FNU:
±
5% of reading (0...100 or 0...1000 NTU/FNU range)
±
2% of reading or
±
0.02 NTU/FNU (0...10 NTU/FNU range)
Resolution
0.0001 NTU/FNU (below 10 NTU/FNU) selectable