5.4 Getting access with activated security access function
The device has a security access function. You can enable and disable this function in the
configuration mode (for more information refer to
Enabling or disabling the security access
on
page 43). With the enabled function the following screen appears any time you press the button
MODE/EXIT:
Execute the following steps to enable and disable the security access
•
Use
or
to select "CONFIG" and press
^
.
•
Use
or
to select "CODE"
•
Use
or
to select "On" or "OFF".
•
If you choose "OFF", press
^
to get back in the normal measuring mode.
•
If you choose "On" first time and press
, the 3 digit security code is set. Using
or
to
select each digit. The flashing digit is the number that is currently being adjusted. Press
to
move on to the next digit.
•
If the code is adjusted, press
to switch to the next item in the menu.
Execute the following steps to get access with an activated security access:
•
Press
or
in the normal measuring mode.
•
The 3 digit security code is set one digit at a time. Press
or
to select the first number. The
flashing digit is the number that is currently being adjusted.
•
Press the button
to accept the first number of the code.
i
Now the second number of the code is flashing.
•
Repeat the previous step with the second number, confirm the choosen number with the
button.
•
Repeat the previous step with the third number and confirm the choosen last number with the
^
button.
i
If you have entered the valid code, the main menu with the arrow beside "CAL" appears. If