6
OPERATION
72
OPTISWIRL 2100
www.krohne.com
04/2020 - 4007600701 - MA OPTISWIRL 2100 R01 en
C
Electronics
Firmware Update Active
A firmware update of the
converter module has been
started.
Wait for firmware update to
finish.
C
Configuration
Sensor Sim. Active
A simulation of the sensor is
active.
Stop the simulation of the
sensor variables.
Current Out. Sim. Active
A simulation of the current
output is active.
Stop the simulation of the
current output variables
HART Sim. Active
Simulation of a measuring
value via HART
®
interface.
Stop the simulation via the
HART
®
interface.
Flow Comp. Sim. Active
A simulation of
measurement value
processing is active.
Stop the simulation of the
measurement value
processing.
Binary Output Sim. Active
A simulation of the binary
output is active.
Stop the simulation of the
binary output variables.
Totalizer Sim. Active
A simulation of the totalizer
is active.
Stop the simulation of the
totalizer variables.
S
Sensor
Flow Meas. Out of Spec.
The vortex frequency or the
volume flow is out of range.
The real flow is higher or
lower than the indicated
value.
Check the process
conditions.
Low Flow Cutoff Undercut
Volume flow is lower than
low flow cutoff (LFC)
Check the low flow cut-off or
the process conditions.
Weak Flow Meas. Signal
The vortex signal is too weak
or disturbed.
Check the process
conditions and the
installation situation.
Eliminate external
disturbances. If the
message recurs, contact the
manufacturer.
Elec. Temp. Out Of Spec.
The electronic temperature
of the sensor module is out
of specification.
Make sure that the device is
operated within the
permissible temperature
range.
S
Electronics
Elec. Temp. Out Of Spec.
The electronic temperature
of the converter is out of
specification.
Make sure that the device is
operated within the
permissible temperature
range.
S
Process
CO Undersaturation
The measurement value is
lower than the lower range
value of the current output.
The current output is
saturated and does not
reflect the effective
measuring value.
Check the process
conditions and the lower
current output limit.
CO Oversaturation
The measurement value is
higher than the upper range
value of the current output.
The current output is
saturated and does not
reflect the effective
measuring value.
Check the process
conditions and the upper
current output limit.
Status
type
Event group
Single event
Description
Actions to eliminate the
event