![KROHNE OPTISWIRL 2100 Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/krohne/optiswirl-2100/optiswirl-2100_handbook_2004298070.webp)
6
OPERATION
70
OPTISWIRL 2100
www.krohne.com
04/2020 - 4007600701 - MA OPTISWIRL 2100 R01 en
Status
type
Event group
Single event
Description
Actions to eliminate the
event
F
Sensor
Sensor Comm. Error
An internal error in the
sensor module has
occurred.
Error in the internal bus
communication or hardware
failure.
For remote version:
Check the sensor
connection.
All versions: Perform a hard
reset. If the message recurs,
contact the manufacturer.
Corrupt Sensor Parameter
Sensor parameters are
inconsistent
Check sensor parameters.
Failed Sensor Input Test
The continuous self test of
the sensor electronic failed.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Failed Sensor MCU Test
The continuous self test of
the sensor electronic failed.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
No Measurement Value
Q
v
or Q
m
have timed out or
are inconsistent.
For remote version:
Check the sensor
connection.
All versions: Perform a hard
reset. If the message recurs,
contact the manufacturer.
Temporary Piezo Error
Piezo signals partly
disturbed.
Check the process
conditions, proper
shielding/grounding and the
installation situation.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Fatal Piezo Error
Piezo signals disturbed.
Check the process
conditions, proper
shielding/grounding and the
installation situation.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Piezo Path disrupted
Sensor electronics error.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Piezo Shortcut
Shortcut on piezo or sensor
electronics detected.
Perform a hard reset. If the
message recurs, contact the
manufacturer.