![KROHNE OPTISWIRL 2100 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/krohne/optiswirl-2100/optiswirl-2100_handbook_2004298017.webp)
DEVICE DESCRIPTION
2
17
OPTISWIRL 2100
www.krohne.com
04/2020 - 4007600701 - MA OPTISWIRL 2100 R01 en
2.2.4 Dual seal
To comply with the requirements of ANSI/ISA 12.27.01 "Requirements for Process Sealing
Between Electrical Systems and Flammable or Combustible Process Fluids", a membrane vent
is integrated in the neck of the device. This vent is located between the primary seal (process)
and the secondary seal (electronics compartment) and works to prevent pressure build-up in the
device neck, thus preventing product from penetrating the electronics compartment in the
unlikely event of a leak in the primary seal.
The seal between the pick-up and the measuring tube is considered as the primary seal.
The material used for this is always the same as that used for the measuring tube itself
(e.g. 1.4435 / 316L for measuring tube made of stainless steel 1.4404 / 316L). When selecting the
material, corrosion resistance depending on process parameters (product, temperature) must
be taken into account.
•
It protects the electronics from the process media.
•
Any leak in the primary seal can be detected.
Even though there is no reason to expect the seal to fail, regular visual checks should still be
carried out to detect any possible leak as early as possible.
In the event of a leak, contact the manufacturer's service department to service or replace the
device.