2
INSTALLATION
16
OPTISONIC 6300
www.krohne.com
06/2019 - 4005857501 - QS OPTISONIC 6300 v2 R01 en
2.8.6 Open feed or discharge
Install the meter on a lowered section of the pipe to ensure a full pipe condition through the
meter.
2.8.7 Position of pump
2.8.8 Position of control valve
Always install control valves downstream of the flowmeter in order to avoid cavitation or
distortion of the flow profile.
Figure 2-15: Open feed or discharge
CAUTION!
Never install the flowmeter at a pump suction side in order to avoid cavitation or flashing in the
flowmeter.
Figure 2-16: Position of pump
Figure 2-17: Position of control valve