4.1 General instructions for programming
After installation of the flow sensor(s) and the electrical connection of the converter, the device
can be switched on and is now ready to be programmed.
Start installation menu
Start installation menu
Start installation menu
Start installation menu
•
Connect converter to power supply and power up converter.
•
Keep left button ">" pressed, until in display appears "release key now".
Figure 4-1: First and second page appear intermittently
CAUTION!
•
If you program the diameter, use the outer diameter of the pipe.
•
For improved accuracy fill in as much details as possible.
•
Fill in the actual transducer distance at menu X7.2.6 (and X8.2.6 if applicable)
•
Run the optimization loop until the transducer distance changes no more than 0.5%.