Operation
30
WARNING! Risk of fire!
Smoking during oxygen therapy is dangerous and is likely to result in facial
burns or death. Do not allow smoking within the same room where the oxygen
concentrator or any oxygen carrying accessories are located!
8.2 Setting the oxygen volume flow
NOTICE!
The maximum limited outlet pressure at the oxygen connector will not exceed 0,7 bar.
The maximum gas output temperature is at most 6 degrees higher than the ambient
temperature.
NOTICE!
The oxygen delivery setting has to be determined for each patient individually with the
configuration of the equipment to be used, including accessories.
1 By turning the dial of the flowmeter,
the volume flow rate can be
adjusted.
by turning the dial clockwise, the flow of
oxygen is decreased;
by turning the dial counter-clockwise,
the flow of oxygen is increased.
NOTE!
The exact flow rate is given at the
center of the ball.
NOTICE!
The flow of the gas to the nasal cannula should be heard or felt in order to check the
function of the device. The hand should be moved back and forth in front of the nasal
cannula. If no gas flows, the connections of the nasal cannula must be checked for leaks.
NOTICE!
If the outlet is closed, the outlet pressure can rise to 0.7 bar. The maximum gas outlet
temperature is 6 degrees above ambient.