aeroplus E
Safety
13
–
Do not lubricate fittings, connections, tubing, or other accessories of the
oxygen concentrator to avoid the risk of fire and burns!
–
Oxygen considerably increases the temperature of a flame and the speed of
combustion!
–
Do not fill the humidifier with flammable fluids!
–
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal
cannula or mask on bed coverings or chair cushions, if the oxygen
concentrator is turned on, but not in use; the oxygen will make the
materials flammable. Turn the oxygen concentrator off when not in use to
prevent oxygen enrichment.
WARNING! Danger of electric current!
Electric energies can cause severe injury. Damaged insulation or components
cause a danger to life.
Therefore:
– Work on the equipment must only be performed by trained professionals.
– Pull the mains plug out before starting work on the device!
– Check mains leads for damage before every use.
ATTENTION! Observe high frequency safety!
Medical equipment can be influenced by (mobile) HF communication equipment
(e.g. mobile phones).
Do not use mobile radio equipment in the vicinity of the aeroplus E.
ATTENTION! Observe electromagnetic compatibility!
Electrical medical equipment is subjected to stringent protective measures
concerning electromagnetic compatibility (EMC) and must be installed and
operated in accordance with the EMC information contained in the
accompanying documents. The following should be noted in particular:
–
Floors should be made of wood or concrete or should be covered with
ceramic tiles. If the floor is covered with a synthetic material, the relative
humidity must be at least 30%.
–
During operation the device must therefore not be exposed to extremely
strong magnetic fields.
–
Magnetic fields at mains frequency must comply with the typical values
found in business or hospital environments.
ATTENTION! Keep the minimum distance!
The air intake of the aeroplus E is located on the back side of the device, the
following is therefore of importance:
–
Place the aeroplus E oxygen concentrator in a well ventilated place
.